Perform meter checkout procedure
Refer to
Troubleshooting Guide
(check inner body operation)
Unplug electrodes and reseat
Fill outer body of electrode with proper amount
of internal filling solution
Remove bubble by redipping electrode
Put electrodes in solution
Fill outer body of electrode with proper amount of internal filling
solution, see
Electrode Preparation
Perform meter checkout procedure (see meter instruction manual)
Ensure that bottom cap is screwed on tight enough to close gap between
bottom cap and body
Check inner body operation
Use recommended carbon dioxide buffer, Orion No. 950210
Check meter and stirrer for grounding
Ensure that membrane is installed properly
Refill outer body of electrode using filling solution shipped with
electrode, Orion No. 950202
Dilute solution
Reduce surface-area-to-volume ratio, slow rate of stirring, avoid
high temperatures
Replace membrane
Allow solutions to come to same temperature
constant temperature before use
Place insulating material between stirrer and beaker