Quick Start Guide

Orion 920Aplus pH/ISE Meter

Calibration of Ion-Selective Electrodes (ISE) with
Two Standards

1.

Verify selected input is currently in use.  If input 1 to be used, the prompt line
should indicate “1-”.  If input not correct, press 

2nd

then 

channel

repeatedly

until selected input is displayed.

2.

Select electrode ID for the electrode in use.  For example to select Fluoride 
(F - ):  press 

2nd

then 

electrode ID

.  Enter 

3

then press 

yes

.  For other

choices press 

2nd

then 

electrode ID

, press 

no

to scroll through other ISE

choice. Enter number of selection and then press 

yes

.

3.

Select “CONCENTRATION” mode by pressing 

mode

until “CON” is

displayed on main display.

4.

Place electrode(s) into first standard and stir moderately.

5.

Press 

calibrate

.  The date and time of the last calibration is displayed.

6.

Press 

2

then 

yes

to select 2 standard calibration.

7.

When “READY ENTER VALUE” is displayed on prompt line, indicating
electrode is stable, enter value of standard and press 

yes

.  The standard value

is entered, stored and the meter display freezes for three seconds.  The meter
automatically switches to standard 2.

8.

Remove electrode(s) from first standard.  Rinse with deionized water.

9.

Place electrode(s) into second standard.

10. When “READY ENTER VALUE” is displayed on the prompt line, indicating

electrode is stable, enter value of standard and press 

yes

.

11. The electrode slope is calculated and then actual value is displayed for 

5 seconds.  Meter then automatically proceeds to MEASURE mode.

12. Remove electrode(s) from standard. Rinse with deionized water. Place

electrode(s) into sample. When “RDY” is displayed, record the sample results.

pH Autocalibration with Two Buffers

1.

Verify selected input is currently in use.  If input 1 to be used, prompt line
should indicate “1-”.  If input not correct, press 

2nd

then 

channel

repeatedly

until selected channel is displayed.

2.

Select electrode ID for the electrode in use.  If pH electrode to be used the
prompt line should read “1-pH”.  If pH not displayed then press 

2nd

then

electrode ID

.  Enter 

1

then press 

yes

.

3.

Select pH mode, by pressing 

mode

until pH displayed on main display.

4.

Place electrode(s) into pH 7 buffer and stir moderately.

5.

Press 

calibrate

.  The date and time of the last calibration is displayed.

6.

Press 

2

then 

yes

to select 2 buffer calibration.

7.

When “RDY” is displayed on prompt line, indicating electrode stability, press

yes

.  The buffer value is entered, stored and meter display freezes for three

seconds.  The meter automatically switches to buffer 2.

8.

Remove electrode(s) from first buffer.  Rinse with deionized water.

9.

Place electrode(s) into second buffer.

10. When “RDY” is displayed on prompt line, press 

yes

.

11. The electrode slope is calculated and then actual value (in percent) is

displayed for five seconds.  Meter will then automatically proceed to the
MEASURE mode.

12. Remove electrode(s) from buffer.  Rinse with deionized water.  Place

electrode(s) into sample.  When “RDY” is displayed, record the 
sample results.