NOTE

:  If using a combination

electrode with a BNC
connector, the reference
electrode input 

(2)

is 

not used.

3.

Attach ATC probe with DIN
connector by sliding the DIN
connectors into the ATC probe
input (

3

) until firmly in place.

pH Measurement
Overview

pH measurement is easy and
accurate with the Orion 920Aplus.  The meter/electrode system is calibrated with
one to five buffers which bracket your sample pH.  For better accuracy, at least a
two buffer calibration is recommended.  

The autocalibration feature automatically recognizes 1.68, 4.01, 7.00, 10.01, and
12.46 buffers.  For best results, an ATC probe is recommended to compensate for
temperature variation.

ISE Measurement Overview

Direct measurement using ion selective electrodes (ISE’s) is fast and convenient
with the Orion 920Aplus.  The meter/electrode system is calibrated with one to
five standards of known concentration which bracket sample concentration.
Unknown sample concentration is then read directly from the display in units used
for calibration molarity, ppm, %, etc. See the appropriate electrode instruction
manual for preparation of sensing and reference electrodes, required solutions
(ionic strength adjusters, standards, etc.)and other special requirements.  For best
results during calibration, always use the most dilute standard first, and always
use a fresh aliquot of standard.

Rec.     Gnd.     ATC/DIN     KF    Input 1   Ref.1     Input 2    Ref. 2

3

1

2

BNC Shorting Cap

Quick Start Guide

Orion 920Aplus pH/ISE Meter

All meter functions are described in detail in the accompanying instruction
manual.  For calibration and measurement only a few steps are required, 
and step-by-step instructions are given in this guide.

Meter Overview

1.

Results and current mode are displayed in the large main field.

2.

Channel, electrode ID, ready, and step-by-step prompts are displayed on the
lower, prompt display.

3.

The keys on the keypad have multiple functions.  Their main function is to
enter numeric values.  They also have primary functions.  These are labeled 
in white and are accessed by pressing 

1st

then the function.  Most keys also

have secondary functions.  These are labeled in green, and are accessed by
pressing 

2nd

then the function.

4.

If you want to escape the sequence you are in  or are not sure of which
function you are in, pressing 

1st

and then 

measure

will take you to the

MEASURE  mode where you can start again.

Meter Connections

Power

1.

Plug line adapter into an appropriate wall outlet.

2.

Plug line adapter firmly into power input on rear panel of meter.

Electrode Connections

1.

Remove BNC shorting cap.

2.

Attach electrode with BNC connector by sliding BNC connector onto
electrode input (

1

), then push down and turn clockwise to lock into position.

Reviews: