2 - 1
2.1 Symptom: Probe 1, 2, 3, or 4 Failure Alarm - SINGLE PROBE FAILURES
Order Possible
Cause
Verification
Action
Probe
Unplugged. Re-seat
connector.
Re-seat all connections.
1
Not fully seated
Verify wires are fully seated in connectors and
connectors are seated together
Replace probe if necessary.
Bad probe
Short.
Open.
Display reads 99.
Verify probe resistance. See
Temperature vs. Resistance Chart
(Page 3-2).
Replace probe.
2
Wrong value.
Verify probe resistance. See
Temperature vs.
Resistance Chart
(Page 3-2).
Replace probe.
3
Micro board.
Temporarily switch suspect probe with control or
heat exchanger probe.
If the same error exists, replace micro board.
2.2 Symptom: Probe 1, 2, 3, and 4 Failure Alarm – (ALL PROBES FAIL)
Order Possible
Cause
Verification
Action
Re-seat connector.
1
Control probe not installed.
Startup only
.
Verify wires are fully seated in connectors and
connectors are seated together.
Reinitialize (re-start) unit.
Reseat all connections.
Replace probe if necessary.
Bad control probe
Display reads 99.
Bad control probe will also
cause a high temperature alarm and lock both
compressors on.
2
Short.
Open.
Wrong value.
Verify probe resistance with
Temperature vs.
Resistance Chart
(page 3-2).
Replace probe.
3
Micro board.
Temporarily switch control probe with heat
exchanger probe.
If the same error exists, replace micro board.
4
Micro board EEPROM
Problem still exists after re-initializing unit.
Re-initialize eeprom in test mode. If problem
remains, replace the micro board. Note: all
calibration, set points and configuration are set to
default.
Summary of Contents for 8600 Series
Page 1: ...7998600 Rev 4 ...
Page 22: ...3 4 ...
Page 23: ...3 5 ...
Page 24: ...3 6 ...
Page 25: ...3 7 ...
Page 26: ...3 8 ...
Page 27: ...3 9 ...
Page 28: ...3 10 ...
Page 29: ...3 11 ...