Temperature; 4.5-50C
Humidity: 0-90%
Atmospheric Pressure: 50 kPa to 106 kPa
ALWAYS
unplug the unit immediately after use and before storing.
ALWAYS
drain the unit before Storage, according to the Preventative Maintenance instructions on
page 13. “Storage” means non-use for over three months.
DO NOT STORE
or store where it can fall into a bathtub, shower, sink, water or other liquid.
Store the unit and pads in a cool, clean, dry space.
AVOID EXPOSURE
to sunlight and high heat conditions for extended periods. Keep away from flames.
When removing and storing the pads or tubing assemblies, replace the caps on the end of the blue tubes.
Dispose of the unit and pads in a manner that is consistent with your federal, state, and local
laws and regulations.
Temperature: 16C-29.4C.
Humidity: 15-90%
Atmospheric Pressure: 70.0 kPa to 106.0 kPa
BEFORE CLEANING:
ALWAYS
power off the unit and unplug the AC / DC power cords from the outlet and the unit before
cleaning. DISCONNECT the brace from the unit. Remove hinges from the brace.
TO CLEAN:
Unit can be cleaned with a neutral disinfectant wipe.
Braces can be cleaned with a neutral disinfectant wipe, soap-water or mild detergent. Rinse brace with
warm tap water and dry with a clean, lint-free cloth. Hinges can be hand washed in cold water with mild
detergent. Allow parts to air dry before re-assembly.
When used in accordance with the cleaning & disinfectant wipe manufacturer’s instructions for use, the
following cleaning & disinfection wipes do not adversely affect the appearance or performance of the
ThermaZone Unit or Brace:
DO NOT:
• use ANY chemical cleaning agents or products containing alcohol on unit, braces or hinges, except as
outlined above.
• wash the unit or hold under running water.
• spray the unit, braces or hinges with cleaning or degreasing sprays.
• scrub the unit, braces or hinges with brushes. DO NOT place unit or braces in autoclave sterilizer.
• attempt to sterilize the unit or braces
• clean the unit, braces or hinges using a laundry machine. Hand wash ONLY.
• dry unit, braces or hinges in machine dryer. Hang or line dry ONLY.
FREQUENCY:
Clean Unit anytime it becomes dirty or between uses with different patients.
Clean Brace anytime it becomes dirty or between uses on different patients.
Clean Hinges anytime they become dirty.
TRANSPORT, STORAGE & DISPOSAL
OPERATION
CLEANING
Cleaning & Disinfectant Wipe
Manufacturer
CaviWipes
TM
Metrex Research
Super Sani Cloth
®
Germicidal Disposable Wipe
Professional Disposables Internation, Inc. (PDI)