Thermal Gas Systems HALOGUARD IR Instruction Manual Download Page 10

Rev.1/26/01

10

CALIBRATION

Haloguard IR monitors are factory calibrated to the specific gas of interest.  Field calibration is not
required for operation in typical machine room environments during the first twelve months of operation
unless a change in gas type is required.

For those who wish to check the factory calibration or change the gas type an optional Span Gas Kit is
available. Instructions for use are included with kit

TESTING

The Haloguard IR

TM

 should be tested at start-up, after any changes to alarm protocol, and after each

alarm level condition has been corrected.

 

During testing, display will indicate TEST MODE

a. Alarm Test –Enter TEST mode with jumper, select alarm with cursor by using

UP/DOWN/ENTER pushbuttons.  Check all alarms sequentially.  Alarm relays, lights, audible
and remote alarms (if equipped) will activate if selected.

b. Reset Test - Push RESET/SILENCE button; alarm relays, audible and remote alarms (if

equipped) will become inactive for 20 minutes.  Display will read RESET and indicate minutes
to completion of reset period.

TROUBLESHOOTING

The most frequent cause of nuisance alarms and fault is improper set-up, primarily due to:
      a.   Failure to install in accordance with instructions.

b. Not allowing sufficient warm-up time.
c. Poor sensor location.

To check Zero Calibration, fill a clean plastic trash bag with clean outside air, tie off with twist-tie, let
sample stabilize inside for 15 - 20 minutes.  Insert sample tube into bag . For models equipped with
scanner allow up to 2.5 minutes per channel for clean air sample analysis
.
Observe LCD response, if displayed PPM value drops lower then Haloguard IR is detecting the presence
of the gas of interest or another gas which absorbs infrared light at the same frequency.

Remove sample tube from bag and observe LCD response. If displayed PPM value moves higher, look
for refrigerant leaks or possible sources of interfering gases  (See “Before Installation). and relocate
sample point.. Keep in mind that the Haloguard IR is much more sensitive than hand held leak detectors,
it may detect a leak when hand held units show no response

We wish to aid with any problems you might experience with the Haloguard IR

.  When contacting

Thermal Gas Systems, Inc. Service Department, please have the following information ready

1. Model Number and Serial Number.
2. Description of the problem.
3. Remove cover. Place jumper on Alarm setting function and record settings.  Place jumper on

Setup function and record settings

4. Advise the condition of all LED’s and the information displayed on the LCD.

Summary of Contents for HALOGUARD IR

Page 1: ...2 R 12 Range 3 R 22 1 0 1000ppm 4 R 134a 5 R 113 6 NH3 7 R 123 O utputs 1 4 x 5A Relays 2 0 5 10 VDC 3 4 20mA 4 RS232 485 5 7 x 10A O ptions Accessories A Audible Alarm B Battery B U UPS E Inlet Outl...

Page 2: ...6 Calibration Testing Trouble Shooting 10 7 Maintenance Specifications 11 FIGURES Page 1 Typical Area Monitoring Installation 3 2 Mounting Dimensions 4 3 LCD Display Information 4 4 Haloguard IR 5 5 A...

Page 3: ...ed near potential leaks in a quiet area or downstream from leak source in area with air movement c Sample tube pick up should be located 18 24 above the floor If using optional scanner locate sample t...

Page 4: ...tdown Alarms are automatically reactivated if alarm condition is not corrected Reset is automatically initiated on start up and after power failure 7 Optional Display Features TWA Integration Haloguar...

Page 5: ...gital Output 4 20mA 0 5VDC RS 232 485 serial Opt Relay Binary Output 4x5A dry contact Std 7x10A dry contact Opt Power In 120 230Vac 50 60Hz Reset Silence Pushbutton Test Pushbutton Sample Outlet Enclo...

Page 6: ...ng Route wire through entry indicated Figure 4 Remove safety cover All connections are made to the terminal block labeled IN 120 250VAC Figure 6 Connect HOT AC wire to Terminal L Connect NEUTRAL wire...

Page 7: ...with power cord provided c On power up RESET and 20 minute count down will appear on LCD For 20 minutes after power up all alarms and relays are automatically disabled Whenever power is disrupted RES...

Page 8: ...cted turns alarm off 5 CAL Install optional activated charcoal filter in channel 1 follow on screen directions to zero calibrate using ENTER pushbutton 6 SPAN Key and instructions for use are included...

Page 9: ...void nuisance alarms remove alarm jumper AUD before beginning To compensate for line losses adjust signal voltage with a digital VOM as follows 1 Open cover on Remote IR expansion module Move jumper f...

Page 10: ...tall in accordance with instructions b Not allowing sufficient warm up time c Poor sensor location To check Zero Calibration fill a clean plastic trash bag with clean outside air tie off with twist ti...

Page 11: ...11 Accessories SPECIFICATIONS Technology Narrow band Photoacoustic Infrared Electrical Supply 115 230 VAC 50 60Hz Output Signal Std 4x3A Dry Contact SPDT Opt 7x10A Dry Contact SPDT Analog Output Opt...

Reviews: