7 Measuring the Humidity Ratio

The Humiport can calculate and 
display humidity or mixing ratio 
in g/Kg or gr/lb (gpp). In order 
to properly calculate this value 
the air pressure or altitude must 
be known. To display the correct 
humidity ratio value, the local 
altitude must be set to within 
±200 ft. of the local altitude.
To adjust altitude, first make sure 
the display on the lower display 
(sensor 2) is set to gr/lb (gpp). 
Now scroll through the lower menu 
items and select the lower right 
arrow on the screen above the Alt. (ft) label on the case. 
Move the wheel up and down to adjust the altitude. You 
can adjust the altitude with 

and 

t

 and confirm by    .

For altitude in your area you can do a web search using 
your city name and the work “elevation” or use the City 
Altitude Guide available from Phoenix Restoration.

8 Changing The Battery

If the symbol “BAT” appears on the display, the batteries 
have to be changed within a few hours. Open the battery 
case on the back of the instrument. Remove the empty 
batteries and insert new ones. Ensure batteries are 
inserted to provide the correct polarity.

Important: The Humiport can be damaged if the batteries 
are inserted incorrectly and reverse polarity results.
Use only EC LR6 AA batteries. Do not uses rechargeable 
batteries. Use only high quality batteries.

Hold  MAX  MIN  AVG

Sensor 1

88:88

-1.8.8.8.8

%rH   °F °C

Sensor 1

-1.8.8.8.8

g/m

3

%rHdp°F °C

Unit1 Unit2 TimeDate
Auto Off CAL 12LH

g/kg   g/lb               Alt(ft)

9 Maintenance And 
Adjustments

Clean the device with a damp cloth as necessary. Do not 
use any cleaning fluids, use just plain water to dampen the 
cloth. Do not touch the sensor. Ensure the unit is off while 
cleaning.
When used for indoor environmental measurement, we 
recommend annual calibration. Recalibration should be 
carried out more frequently in a harsh environment.
Recalibration should only be performed by a calibration 
laboratory, or by using a standard calibration method 
such as use of saturated salt solutions. Regardless of 
the method, calibrations should occur at a minimum of 
two %RH concentrations. Humidity sensor can also be 
calibrated to another recently calibrated humidity sensor.
Phoenix Restoration (Therma-Stor LLC) offers calibration 
services for all types and makes of hand held temperature 
and humidity monitors. For more information on the 
Phoenix Calibration service call 1-800-533-7553.

10 Specifications

Measurement Range:

 

RH 

0 to 95% RH

 

Temp 

-4 to 122°F (-20 to 50°C)

Calculated Values:

 

Dew Point 

°F, °C

 

Humidity Ratio 

grams/Kg dry air,  

 

 

grains/lb dry air (gpp)

Accuracy:

 

RH 

±2% RH

 

Temp 

±0.5°F at 32 to 103°F  

 

 

otherwise ±0.9°F

 

 

(±0.3°C at 0 to 40°C  

 

 

otherwise ±0.5°C)

Resolution:

 

RH 

0.1%

 

Temp 

0.1°F, (o0.1°C)

Response Time:

 

RH 

T90 <6.0 seconds

General:

 

Supply Voltage 

4-1.5V Battery

 

Battery Life 

500 hours (typical)

 

Working Temp Range 

32 to 122°F (0 to 50°C)

 

Storage Temp Range 

-4 to 140°F (-20 to 60°C)

 

CE Compatiblility 

EN61000-6-2, EN50147-3

Weight:

 

0.5 lbs. (200 g)

Summary of Contents for Humiport 10

Page 1: ...r head or near the floor Each Humiport Meter offers a large display for easy reading and all of the important restoration information including temperature either F or C relative humidity dew point F...

Page 2: ...the Humidity Ratio 3 7 Replacing the Battery 3 8 Maintenance and Adjustment 3 9 Specifications 4 10 Warranty Sensor Repair 5 Serial No ___________________________ Purchase Date ______ ______ _____ Dea...

Page 3: ...ivate selection with or interrupt with t or by not pressing the wheel for 20 seconds 5 Upper Menu The following functions can be selected in the upper menu HOLD MAX MIN AVG Once the upper menu is acti...

Page 4: ...only be performed by a reputable calibration laboratory The original factory calibration can be reset by adjusting the offset for CA2L and CAL2H to 0 0 RH 00 0 rH 80 0 rH SENSOR 1 20 0 F C SENSOR 2 80...

Page 5: ...VG Sensor 1 88 88 1 8 8 8 8 rH F C Sensor 1 1 8 8 8 8 g m3 rHdp F C Unit1 Unit2 TimeDate Auto Off CAL 12LH g kg g lb Alt ft 9 Maintenance And Adjustments Clean the device with a damp cloth as necessar...

Page 6: ...t If you believe your Humiport has a defect in workman ship contact Phoenix Restoration to arrange for a replacement unit If you have damaged your Humiport and you feel it may need repair or replaceme...

Page 7: ...enix if the unit is found to be defective according to the warranty it will be repaired or replaced at no charge If the product is found not to be covered by the warranty the customer will be offered...

Reviews: