pH Wireless Ver.21-7 12
5. Radio chanel
pH Configuration
6 Radio channel
pH 7.40 reg 20%
TX 30 canal 07
This function allows you to check transmission status and change the transmission
channel.
pH 7.40
pH reading value
Reg 20%
Calculated dosage regulation
TX30
Command sent to the pump. When transmission is
taking place this parameter flashes every second
Canal 07
Current transmission channel
If you have interference problems with other equipment that may be installed in the
vicinity you can choose a different channel:
1) In Radio channel menu, choose a different channel with the
arrows
2) Press and hold the
prime button
(3) on the pump
3) Press
OK
The communication channel will change at the chlorinator and at the pump. The blue
pilot of the pump will flash again indicating that it is receiving data from the chlorinator.
6. Radio channel
pH Configuration
7 Dosing delay
Delay
OFF
Time
20 min
The probes take a certain time to generate a valid response after being disconnect-
ed. This function allows you to set a delay after turning on the device during which
the dosage remains on hold, thus avoiding an incorrect dosage. We recommend to
always activate a delay time of at least 15 min. When dosing is on hold, after turning
on the instrument, the main display will show dashes instead of pH value: pH—-.—-.
Place the cursor (
MENU
) under the word “OFF” and use
arrows
to activate (ON) or
deactivate (OFF) the dosing delay. Place the cursor under the word “min” and use
arrows
to set delay time in minutes. Press
OK
to save and exit.