| 30 |
Notes on electromagnetic
compatibility
Electrical leads, lead lengths and
accessories
The THERA-fit may only be operated with the
original power cord.
Waning note about siting
The apparatus or system may not be placed
in the immediate proximity of or stacked to-
gether with other equipment. If operation in
the immediate proximity of or stacked to-
gether with other equipment is absolutely un-
avoidable, the apparatus or system should be
observed in order to make sure it is operating
as intended in that location.
Compliance levels
The apparatus complies with the interference
immunity testing levels required by
IEC 60601.
Interference emission
Guidance and manufacturer’s declaration on electromagnetic emissions
The THERA-fit is intended for operation in the electromagnetic environment specified below.
The THERA-fit owner or user should ensure that the apparatus is used in such an environment.
Emission measurements
Compliance
Guidance on electromagnetic envi-
ronment
High-frequency emissions
as per CISPR 11
Group 1
The THERA-fit only uses high-frequency
energy for its internal functions. There-
fore, its high-frequency emissions are
very low and it is unlikely that adjacent
electronic equipment will suffer interfer-
ence.
High-frequency emissions
as per CISPR 11
Class B
The THERA-fit is intended for use in all
types of facility including:
- residential areas
- facilities that are directly connected to a
public power grid that also supplies resi-
dential buildings.
High-frequency oscillations
as per IEC 61000-3-2
Class A
Voltage fluctuations/flicker
as per IEC 61000-3-3
complies