The Cary LED At-A-Glance Manual Download Page 2

Assembly Instructions for the LED At A Glance Remote Display 

The Cary Company

1195 W. Fullerton Ave.

Addison, IL 60101

thecarycompany.com

Contact us for help: 

Ph: 630-629-6600

1. Unscrew mechanical gauge vial retainer and lift vial and retainer off.

2. Place magnet holder on indicator, and replace vial and retainer over magnet holder and screw 

retainer back down.

3.

Skip step one and two if your gauge and LED were ordered together.  In this case, your indicator 

and magnet

 

are already installed.

4. The clamp, clamp screws, and transmitter should already be assembled.  Slide the transmitter and 

clamp down over the vial.

5. Looking at the top view (the picture on bottom right), line the transmitter up with the magnet as 

shown in the picture and tighten the two lock down screws.

6. To install battery, unscrew the battery box and pull the tab for access to battery compartment.

7. Remember that this is the location of the alarm settings (see user guide).

8. View the current fuel level using the mechanical float gauge.

9. Press the “Push” button once to check the level on the remote display.

10.Slide the transmitter up or down until the mechanical gauge reading matches the electronic gauge 

reading.

11.Tighten the lock down screws to lock the transmitter into place.

12.Mount the display in a convenient location. 

Indicator 

Magnet Holder 

Retainer 

Transmitter 

Vial 

Clamp 

Lock Down Screws 

Clamp Screws 

Top View 

Before installing battery, see user guide for alarm settings.  Default settings have the alarms turned off. 

Retainer 

Clamp 

Vial 

Magnet 

Transmitter 

Reviews: