8 Test the device
Caution!
Do the following tests once per year or after repair.
• Electrical safety test, see
on page 40
• System operation test, see
on page 40
• Overtemperature alarm test, see
on page 41
8.1 Electrical safety test
8.1.1 Necessary items
• the device
• IV pole (optional)
• saline or other crystalloid fluids
• standard set
• syringe or hospital administration set
• medical electrical safety tester (refer to IEC 60601-1)
• metal catheter
8.1.2 Preparation
1.
Connect a metal catheter to the end of the patient line.
2.
Open the clamps on the hospital administration set and the patient line.
3.
Make sure that the complete standard set and the catheter are free of air.
4.
Close the clamps of the hospital administration set.
5.
Connect the metal part of the catheter to the safety tester. This is an applied part on the
terminal of the safety tester.
8.1.3 Procedure
1.
Do the electrical safety test, refer to IEC 60601-1, for a class II, Body Floating Device.
8.2 System operation test
8.2.1 Necessary items
• the device
• IV pole (optional)
• saline or other crystalloid fluids 21 ± 3°C
• standard set
• hospital administration set
• measuring cup and timing device to measure the flow rate
• thermometer to measure the outlet temperature