The graph screen will display the following data over time (hours and mins) for the
pressure of the cell (bar), the output pressure (bar), the flow rate of hydrogen
(NmL/min) out of the system, and the consumed hydrogen (NmL).
5.5
Using and Changing a ScavCart™
In order to maintain water purity at an acceptable level for the H-Genie
®
and to ensure
a good hydrogen cell longevity, then it is mandatory to use a ScavCart™. The
ScavCart™ should be changed every 2 months. ScavCart™s can be ordered directly
from ThalesNano. Contact
Do not use your own resin as a water purity scavenger. Only use official
ThalesNano certified products. Failure to do so will invalidate product warranty.
A special resin with a
particle size is used in ThalesNano ScavCart™s. Only use
a
ScavCart™ for a maximum of two months. Using ScavCart™s beyond this
timeframe may cause leaching of the resin into the water tank.
Figure 24: How to attach the metal cap to the
ScavCart™ and
how to install it in the H-Genie
®
To use the ScavCart
™, take the plastic holder and screw it into the metal cap. Place
the ScavCart
™ into the H-Genie
®
water tank. Please ensure water tank is full every
day before use.
Once the ScavCart
™ has expired, then dispose in solid waste or ship back to
ThalesNano for recycling. Address is given at the end of the manual.
5.6
Turning off and shutting down the H-Genie
®
To turn off/shutdown the H-Genie
®
. Press the Shutdown button on the front of the
system.