Enter
Brightness
Mode
Step 1
Step 2
Step 3
Operational
Display
Press
2
Press
at same
time
Press
at same
time
Step 4
Step 5
OR
OR
Press
1
Press
1
Press
1
TWO-POINT
CALIBRATION
Step 7
Step 8
Step 9
Step 10
Step 11
Step 6
OR
Operational Display
Press
1
From Step 6
To Step 7
Pass Brightness Mode and
enter Calibration Mode
Select the No Function
Calibration Mode [000]
Save Calibration Mode [000]
setting and enter Code 1
Enter Cal Mode [111]
For 2-point calibration
of CH1
5.2. Apply the LOW
input signal
5.1. Adjust
display to
desired reading for
zero input
7.2. Apply the HIGH
input signal
7.1. Adjust display to
desired reading for
span input
Set reading for zero load into
meter and enter Span Mode
Save zero and span settings
and re-enter Calibration Mode
LOW
Signal
HIGH
Signal
2-POINT CALIBRATION
Set Calibration Mode to [111]:
1st Digit = 1
Selects calibration procedures
2nd Digit = 1
Selects 2-point calibration
3rd Digit = 1
Selects CH1 for calibration
Example
Example
OR
Press
at same
time
Press
at same
time
Exit
Code2.
Return to
Operational
Display
LOW
Signal
HIGH
Signal
The low input source is applied to
the meter when setting the zero
value.
The high input source is applied
to the meter when setting the
span value.
Two-point Calibration
Two-point calibration is the most commonly used method of
calibrating Tiger 320 Series meters when a low and high input
source is available.
Example Calibration Procedure
Calibrate channel 1 (CH1) using the two-point calibration meth-
od. Set the calibration mode display to
[111]
.
START HERE