FL-B101D40PF Power factor (D0109) Manual

Page 5

From Two Point

Digital Calibration 

Mode See Page 4

STEP A

STEP B

STEP C

STEP D

To Bargraph Center Point

Display Mode Selection

DIGITAL SPAN SELECTION

FOR BARGRAPH DISPLAY

[X•XXX]

[XX•XX]

[XXX•X]

[XXXX•]

[XXXX]

DECIMAL POINT AND

BRIGHTNESS SELECTION

Bargraph Display

Scale Lo Range

Setting [bLo]

Adjust to: 50

Bargraph Display

Scale hi Range

Setting [bhi]

Adjust to: 150

Set the Decimal

Point [dp]

Select 

[XX•XX]

 

STEP A  Enter the Calibration Sub Menu Mode

 

1)  Press the 

P

 and   buttons at the same time. Display toggles 

between [CAL] and [oFF]. 

 

2)  Press  the 

P

  button.  Display  toggles  between  [bhi]  and  the 

p re v i o u s  setting.

STEP B  Set the Digital Span of the Bargraph Display

 

1)  Using the   and   buttons, adjust the display to[150]  

high parameter reading.

 

2)  Press the 

P

 button. Display toggles between [bLo] and the 

previous setting.

 

3)  Using  the    and    buttons,  adjust  the  display  to  [50]  low 

parameter reading.

 

4)  Press the 

P

 button. Display changes from [50] to [dP].  

STEP C  Set the Decimal Point

 

1)  Using the   and   buttons, adjust the display to the   

  

     

[XX.XX]

 

decimal point setting.

 

2)  Press the 

P

 button. Display toggles between [br] and the 

 

  previous brightness setting.

STEP D  Set the Bargraph and Digital Display Brightness

 

1)  Using the   and   buttons, adjust the display to the 

 

     desired brightness setting (4 is the brightest setting).

 

2)  Press the 

P

 button. Display toggles between [Cto] and [oFF]

 

.

Digital Span Selection For Bargraph Display

Decimal Point and Brightness Selection

Bargraph Center Point Display Mode Selection

STEP G  Bargraph Center Point Mode Selection (See example above)

 

1)  To select bargraph center point mode, press the   or 

  

  button. Display changes from [oFF] to [on].

 

2)  Press the 

P

 button. Display toggles between [diSP] and  

 

     [on] or [oFF].

STEP H  Digital Display ON/OFF Selection

 

1)  To set the display to [oFF], press the    or   button. 

 

     Display toggles between [diSP] and [oFF].

 

2)  Press the 

P

 button. The display exits the calibration mode  

  and  returns  to  the  operational  display.  Only  the  bargraph 

 

  display is on and the digital display is off.

If the digital display is selected to be off, pressing any button to 

make programming changes or to view setpoints activates the digital display. 

When the procedure is complete, the digital display will then automatically 

switch off.

The Display/Bargraph settings are now complete. 

Operational Display

STEP G

STEP H

From Decimal Point

and Brightness Selection

BARGRAPH CENTER POINT

DISPLAY MODE SELECTION

Bargraph Center Point

Display Selection [cto]

Select [on]

Summary of Contents for Leopard Series

Page 1: ...or Bargraph Indicator Bargraph center zero function Digital display blanking Four level brightness control accessed by the button and adjusted by the button Software Features Bargraph Center Point Display Mode Selection 5 Case Dimensions 9 Component Layout 7 Connector Pinouts 6 Connectors 6 Controls and Indicators 2 Decimal Point and Brightness Selection 5 Software Features 1 Software Logic Tree 3...

Page 2: ...at an alternative logic branch will be followed when a particular option is present When two buttons are shown side by side and enclosed by a dotted line they must be pressed at the same time then released to go onto the next programming step If an X appears through a digit it means that any number displayed in that digit is not rel evant to the function being explained P Span 10000 P P When the a...

Page 3: ...ut Lo cLo will occur This branch will only appear if the analog output option is installed DIGITAL SPAN SELECTION FOR ANALOG RANGE OUTPUT See Page 6 DIGITAL DISPLAY ON OFF See Page 5 BARGRAPH CENTER POINT DISPLAY MODE SELECTION See Page 6 TWO POINT ANALOG OUTPUT RANGE SETTING AND CALIBRATION See Page 7 Set the Decimal Point dp Select XX XX Bargraph Display Scale Lo Range Setting bLo Adjust to 50 B...

Page 4: ...Digital Span Selection for Bargraph Display and the display flashes bhi and the previous setting ERROR Indicates Unsuccessful Calibration If the calibration was unsuccessful the display indicates Err the new calibration settings just entered will not take effect and the previously stored setting will remain The three most likely causes of an error during calibration are 1 The full scale and zero s...

Page 5: ...ous brightness setting STEP D Set the Bargraph and Digital Display Brightness 1 Using the and buttons adjust the display to the desired brightness setting 4 is the brightest setting 2 Press the P button Display toggles between Cto and oFF Digital Span Selection For Bargraph Display Decimal Point and Brightness Selection Bargraph Center Point Display Mode Selection STEP G Bargraph Center Point Mode...

Page 6: ...a unique plug and socket outline to prevent cross connection The main board uses standard right angled connectors Replacement 2 3 and 4 pin plug connectors are available see Accessories on page 10 Input Signal Pins 1 to 6 Pins 1 to 6 are reserved for the input signal conditioner See the data sheet for the selected input signal conditioner Pins 17 to 21 Rear Panel Switches Pin 19 Programming LOCK B...

Page 7: ...ODULE 24V Exc IP07 Universal Process Input 2V 5V 10V 20V 200V 2mA 20mA Custom Custom 20mA 2mA 200V 20V 10V 5V 2V VOLTAGE CURRENT UNIVERSALPROCESS Processinput Requires Digital Calibration HI LO OFF ON 24VEXC 224C Input Range Header IP07 24V Exc on off Header Signal Type Header Voltage Current ...

Page 8: ...l adapter enables modern DIN meters to fit in existing cutouts individually or stacked when replacing old 6 edgewise mechanical pointer meters Fits existing cut outs for 6 150 mm edgewise switchboard pointer meters from Crompton G E Westinghouse Yokogawa and most others Width 43 7 mm to 48 mm 1 72 to 1 89 Height 143 4 mm to 149 mm 5 62 to 5 88 Fits 6 Edgewise Pointer Meter Cut Outs Panel Adapter W...

Page 9: ... IO board The adapter snaps on the 36x144 mm 1 42 x5 69 case and enables single unit or stack mounting in an existing 6 edgewise pointer meter cut out 100 0 25 50 75 100 0 25 50 75 SP1 SP2 SP3 SP4 Zero Span Zero Span Panel adaptor plates are available to retrofit most existing panel cutouts 31mm 1 22 These dimensions are increased by 2mm 0 08 when the metal surround case is installed For extra str...

Page 10: ...odel number FL B101D40PF 144x36mm 101 Segment Bargraph 4 Digit Scale Factor Standard Options for this Model Number Order Code Suffix Description Display VRR Red LED Bargraph w 4 Digit Red DPM Vertical VGG Green LED Bargraph w 4 Digit Green DPM Vertical VGR Green LED Bargraph w 4 Digit Red DPM Vertical VRG Red LED Bargraph w 4 Digit Green DPM Vertical HRR Red LED Bargraph w 4 Digit Red DPM Horizont...

Reviews: