2-1-2. CR Mode
Description
In Constant Resistance Mode, the unit will maintain a
constant resistive load by varying the current.
CR mode uses
Ω (resistance) or S (conductance) for the
setting units.
For more details on CR mode,
please see the appendix on page 172 (
7-5-2.CR Mode
).
Warning
If you change the mode or the range when the load is
already on, the load will be turned off automatically.
Operation
1.
Make sure the load is off.
2.
Press
Main
.
3.
Select CR mode with the
Mode
[F1]
soft-key.
4.
Select the current range with the
I Range
[F2] soft-
key.
I Range:
High, Middle, Low
5.
Select the voltage range with the
V Range
[F3] soft-
key.
V Range:
High, Low
6.
Set the resistance or conductance level parameters
using the scroll wheel and number pad.
For Static mode, set
CR A Value
and/or
CR B Value
.
For Dynamic mode, set
Level1
and
Level2
.
The maximum and minimum conductance/ resistance
levels depend on the selected current range.
7.
To add CV mode to CR mode (CR+CV),
see page 42 (
2-1-6. +CV Mode
).
8.
Set the remaining basic configuration settings such as
the slew rate, and switching function settings.
See page 47 (
2-2. Basic Configuration
) for details.
Display
Mode
CR
I Range
H 8.75A
V Range
L 80V
Function
Static
Configure
0.000
V
0.0000
A
0.00
W
SlewRate
14.000
CR B Value
0.54422
CR A Value
0.54422
LOAD
RS232
A Value
Fine
01/Oct/2017
mA/uS
Ω
Ω
Mode
Current range
Voltage
range
Active
setting
Conductance/
Resistance
settings
Summary of Contents for LSG-1050H
Page 24: ...14 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6...
Page 185: ...175 7 6 Operating Area 7 6 1 LSG 175H H Range M Range L Range...
Page 186: ...176 7 6 2 LSG 350H H Range M Range L Range...
Page 187: ...177 7 6 3 LSG 1050H H Range M Range L Range...
Page 188: ...178 7 6 4 LSG 2100H H Range M Range L Range...
Page 204: ......