3-5-1. OPP Test function setting parameters
Parameters
No.
Selects one of 12 OPP test setup memories.
Memo
A user-created note for the currently selected
OPP function.
Range
Select the Range of CP Mode.
(High, Middle, Low)
Start Watt
(Start W)
Starting start watt value for the test.
End Watt
(End W)
The watt value that will end the test.
The value must be higher than the OPP
value of the power supply you are testing.
This parameter is used as a fail-safe for if the
over power protection of the power supply
fails.
If the measured watt is reaches End Watt
value it would then indicate that the power
supply OPP failed.
Step Watt
(Step W)
Sets the step resolution of the watt.
Step Time
(Step T)
Sets the execution time of each step. (10ms
to 50s)
Trig Delay
Time
(Delay)
Sets a delay corresponding to the time a Trig
Voltage can be expected after each step
Watt is applied (the delay time must be less
than the Step time).
Trig Voltage
(Trig V)
Sets the trigger to a level needed to see
when the power supply OPP has been
triggered.
When the power supply OPP has been
triggered, its voltage output will reset.
The voltage trigger level is used to test to
see if the voltage output has been reset.
Last Watt
(last W)
Sets the final watt value after OPP has been
tripped.
This is the steady-state watt draw after the
OPP has been tripped.
Note
This mode can only be used under CP mode.
Summary of Contents for LSG-1050H
Page 24: ...14 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6...
Page 185: ...175 7 6 Operating Area 7 6 1 LSG 175H H Range M Range L Range...
Page 186: ...176 7 6 2 LSG 350H H Range M Range L Range...
Page 187: ...177 7 6 3 LSG 1050H H Range M Range L Range...
Page 188: ...178 7 6 4 LSG 2100H H Range M Range L Range...
Page 204: ......