46
Front Panel Operation
1.
Configure the thermocouple type and unit. Pressing CONFIG > SHIFT > TEMP (TCOUPL),
and then using PREV and NEXT to select the type and unit.
2.
Measure a known and accurate reference temperature (T1). The result showing on the
DL-2060 is (T2).
3.
Adjust the simulated value (23.0°C to 23.0°C + T3.)
4.
Configure the Simulated value by pressing CONFIG > SHIFT > TEMP(TCOUPL) > NEXT >
SIMULATED > ENTER, and by buttons
「
PREV
」
or
「
NEXT
」
and
△
&
▽
assistance to
adjust the value.
Remote Interface Operation
Use the following commands to set up the thermocouple configuration.
SENSe:UNIT {Cel|Far|K}
SENSe:UNIT?
SENSe:TCOuple:TYPE {E|J|K|N|R|S|T}
※
SENSe:TCOuple:RJUNCtion:SIMulated {<
value
>|MINimum|MAXimum}
※
SENSe:TCOuple:RJUNCtion:SIMulated?
※
This command is available with firmware version 1.05 or upward.
3.1.9 Remote Interface Selection
The multimeter supports GP-IB (G/VG type), RS-232C(R/VR type) and USB interfaces,
but only one interface can be activated at a time. If you are using GP-IB, you must set
the address for the multimeter. You can set the address to any value from 0 and 31. The
address is set to “22” when the multimeter is shipped from the factory.
If you are using RS-232C, you must set the baudrate and parity for the multimeter.
※
Note: The remote interface can only be set through the front panel operations.
How to select a remote interface
Press MENU and then use
◁
and
▷
to select INTERFACE. Then Press ENTER on your
choice of USB or GPIB interface.
How to set address when using GP-IB
Press MENU and then use
◁
and
▷
to access INTERFACE. Select GPIB to set address.
Use
◁,
▷,
△
and
▽
to adjust the numbers to the desired address. Press ENTER to set
it.
How to set baudrate when using RS-232C
Press MENU and then use
◁
and
▷
to access INTERFACE. Select RS232, Select