Required Equipment
2-2
2.1
Required Equipment
-
One dual-output dc power supply (
±
5 V, 1 A output minimum)
-
Two dc current meters with resolution to 1 mA and capable of the
maximum current the dc power supply can supply.
-
50-
Ω
source impedance function generator (1 MHz, 10 V
PP
sine wave)
-
Oscilloscope (50-MHz bandwidth minimum, 50-
Ω
input impedance)
-
3 BNC-to-SMA cables
-
BNC-to-BNC cable
-
5 Banana-to-Banana wires; 4 red, 1 black
2.2
Power Supply Connection (Refer to Figure 2−1)
1) Set the dual dc power supply to
±
5 V. If available, set the current limit on
the dc power supply to 100 mA.
2) Make sure the dual dc power supply is turned off before proceeding to the
next step.
3) Connect the positive (+) terminal of the power supply to the positive (+)
terminal of the current meter number 1.
4) Connect the negative (–) terminal of the current meter number 1 to +VS
(J8).
5) Connect the common ground terminal of the power supply to GND (J7).
6) Connect the negative (–) terminal of the power supply to the negative (–)
terminal of the second current meter.
7) Connect the positive (+) terminal of the current meter number 2 to –VS
(J5).
Figure 2−1. Power Supply Connection for
±
5 Vdc
Summary of Contents for THS4503
Page 1: ...THS4503EVM June 2002 HPL User s Guide SLOU132...
Page 8: ...vi...
Page 14: ...1 4...
Page 18: ...2 4...
Page 27: ...5 3 EVM Hardware Description Figure 5 2 Bottom Layer 2 Ground and Signal...