The EVM requires a 7-V to 28-V input applied to PWR_IN and GND_IN. The power supply should be
current limited to 100 mA. The default jumper settings connect the resistive bridge emulator to the VBRG
outputs and to the VINP inputs of the PGAxxx device and are described in full in
Table 1. Default Jumper Settings
Jumper
Setting
Function
J4
Pins 1-2 Closed
Connect VINTP to RT1 10-k
Ω
thermistor
J5
Pins 1-2 Closed
Connect top of resistive bridge emulator to VBRGP
J6
Pins 1-2 Closed
Connect resistive bridge emulator output to VINPP
J7
Pins 1-2 Closed
Connect resistive bridge emulator output to VINPN
J9
Pins 2-3 Closed
Power 3.3-V LDO from VCC
J11
Open
Disconnect SENT output
Table 2. Additional Jumper Settings Description
Jumper
Setting
Function
J4
Pins 2-3 closed
Connect DAC_CAP to RT1 10-k
Ω
thermistor
J5
Pins 2-3 closed
Connect top of resistive bridge emulator to TEST1
Open
Connect VBRGP to J3 pin 1
J6
Pins 2-3 closed
Connect resistive bridge emulator output to VINTP
Open
Connect VINPP to J2 pin 1
J7
Pins 2-3 Closed
Connect resistive bridge emulator output to VINTN
Open
Connect VINPN to J2 pin 2
J9
Pins 1-2 Closed
Power 3.3-V LDO from USB VBUS
J11
Closed
Connect SENT output
3
EVM Setup and Operation
3.1
Quick Start Procedure
The PGAxxxEVM-037 is designed to be ready for immediate evaluation out of the box with the onboard
resistive bridge emulator as a simulation of a pressure sensor input. The following procedure details the
steps necessary to begin evaluation with the PGAxxxEVM-037 and the PGAxxx EVM GUI.
1. Collect the following supplies:
•
PGAxxxEVM-037
•
PC running Windows 7 or later
•
Micro-USB to USB cable
•
A single power supply unit, battery, or AC/DC adapter to provide a voltage output from 7 V to 28 V
and a current output of up to 100 mA.
2. Follow the instructions in the
Installation
section (see
) of the User's Guide to install and run
the PGAxxx EVM GUI.
3. With the power supply off, connect the positive output of the supply to the PWR_IN connector on the
EVM, and the power supply ground to the GND_IN connector. Only turn on the power supply output
after connecting to the EVM.
4. Plug the micro-USB cable into the USB port on the EVM, and the USB side of the cable into the PC.
The EVM setup should look similar to the one shown in