Texas Instruments MSP-SA430-SUB1GHZ Quick Start Manual Download Page 2

 

 

1. 

Software and Driver Installation 

Run the installer from attached resource CD or go to 

www.ti.com/sa430

 

for latest version. This will install the Spectrum Analyzer Graphical User 
Interface 

(GUI), the full User’s Guide and the required device drivers.  

 

2. 

Connect  the Hardware 

Use included USB cable to connect the SA430 device to a spare USB 
port on your computer. The Microsoft® certified driver is automatically 
detected and installs hardware as  

MSP-SA430-SUB1GHZ 

– CDC 

virtual COM port device. When the SA430 is connected to the PC, the 
SA430 blinks a green LED followed by a steady red LED. The SA430 is 
now ready to be used. 

 

3. 

Connect  RF 

The SA430 is equipped with a standard 50 Ohm SMA connector. Use 
the included SMA antenna for the first measurements. See User’s Guide 
for details and effects of the antenna. 

 

CAUTION: 

Absolute maximum input level is +0 dBm. Exceeding 

this level can damage the SA430 hardware. 

 

4. 

Launch Spectrum Analyzer Software 

By default, setup.exe creates a shortcut to the software in the 
Windows® Start Menu and on the Desktop from where it can easily be 
launched. It can also be started by running 
%sa430_install_path%\SA430GUI.exe. 
 
The SA430 GUI starts in the Hardware tab to allow selection of the 
desired hardware.  Assuming only one SA430 connected to this 
computer the connection can easily be made by pressing the Connect 
symbol or the Connect button in the hardware tab. 

     

 

 
 
 
 
 

 

5. 

Set Frequency Range of Interest  

Select the desired frequency range (one of the three supported 
ranges), then specify the center frequency and span to measure. You 
can either use Center/Span or Start/Stop frequencies 

– both settings 

are equivalent. 

 

6. 

Set expected Amplitude Level  

Select the Reference Level (RefLvL) accordingly to the expected input 
level. The Reference level gives the maximum level that can be 
detected without saturation of the input stage. If signal strength is not 
known, it is a good idea to start with a low reference level and 
increase it if high readings or artifacts are seen. 

 

7. 

Start Measurement  

Once all settings are made the measurement is triggered with the 
start button. Use this button whenever you want changed RF settings 
to become active. 

 

8. 

First Output  

In the left Graph Window the measured spectrum will be displayed. It 
gives power level information over frequency. By default the actual 
trace is shown, as well as the maximum value measured since the 
start button was pressed. More traces can be added from the Trace 
tab on the right. 

 

9. 

Marker 

Markers are available to allow for easier reading of measurement 
results. They are enabled in the Marker tab by selecting a trace. With 
the jog dial the markers can be moved in frequency. The jog dial is 
assigned to the marker associated with the button being pressed. 

 

10. 

Further information

 

More information on how to use the Spectrum Analyzer can be found 
in the Spectrum Analyzers User’s Guide. You can easily launch it from 
within the GUI in the Help menu or at 

www.ti.com/sa430

.  