3.2.2.3.4 Alarms Tab
The
Alarm
tab on the
Quick-Start
page is a special medical alarm function provided by the DAC53701. The
settings for this function are as shown in
. The ramp generation settings are similar to that of function
generation: code-step, slew-rate, margin-low, and margin-high.
Figure 3-10. DAC Quick-Start Page: Alarms Tab
In addition to these settings, the medical alarms have three more settings: interburst time, pulse-off time, and
pulse-on time. These settings have four common settings that take different values depending on the type of
alarm selected. Refer to the
for these timing details. There are three types of alarms: low-
priority, medium-priority, and high-priority. The radio buttons are used to trigger any one of the alarms. The
register settings are programmed or retrieved using the EEPROM
PROGRAM
or
RELOAD
buttons, respectively.
Detailed Description
20
DAC53701EVM
SLAU841 – OCTOBER 2020
Copyright © 2020 Texas Instruments Incorporated