3 Interface Adapter
The BQStudio software uses either the on-board USB interface adapter or separate EV2400 interface board to
provide communication with the EVM board from the computer. The on-board interface identifies as an EV2400.
The EV2400 uses operating system drivers and no separate installation is required.
When connecting an external EV2400 change the J7 jumpers on the board from the
INT
to the
EXT
position.
Since the external EV2400 provides pull ups, do not install shunts at the
PU
positions.
shows an
example of the EVM connected with an external EV2400.
DC Power Supply
+ -
EV2400
I2C
Install cell
simulator shunts
Position shunts to “EXT”
Figure 3-1. EVM Connection with External EV2400
EV2400 versions 0.28 and newer are compatible with the on-board interface. The latest version is
recommended. If updates are available for the on-board interface adapter, they are available from the EV2400
tool folder on the web,
. Do not install the EV2400 firmware version 0.18 in the on-board
adapter.
The BQ76922 devices are not supported with the older EV2300 interface.
CAUTION
Do not program the on-board adapter with version 0.18 or earlier firmware.
Interface Adapter
6
BQ76922EVM Evaluation Module
SLVU957A – SEPTEMBER 2019 – REVISED NOVEMBER 2021
Copyright © 2021 Texas Instruments Incorporated