Operation
8
SLUUBF9 – March 2016
Copyright © 2016, Texas Instruments Incorporated
bq4050EVM-561 Evaluation Module
3.3
Calibration
The bq4050EVM must be calibrated to ensure accurate value reporting. This can be done by going to the
Calibration
window in bqStudio (
Figure 5. Calibration Screen
3.3.1
Voltage Calibration
The following list provides information about voltage calibration:
•
Measure the voltage from Cell 1 to 1N and enter this value in the
Applied Cell 1 Voltage
field and
select the
Calibrate Voltage
box.
•
Measure the voltage from BAT+ to BAT– and enter this value in the
Applied Battery Voltage
field and
select the
Calibrate Battery Voltage
box.
•
Measure the voltage from PACK+ to PACK– and enter this value in the
Applied Pack Voltage
field and
select the
Calibrate Pack Voltage
box. If the voltage is not present, turn the charge and discharge
FETs on by pressing the
FET_EN
button on the
Commands
panel.
•
Press the
Calibrate Gas Gauge
button to calibrate the voltage measurement system.
•
Deselect the
Calibrate Voltage
boxes after voltage calibration has completed.
3.3.2
Temperature Calibration
The following list provides information about temperature calibration:
•
Enter the room temperature in each of the
Applied Temperature
fields and select the
Calibrate
box for
each thermistor to be calibrated. The temperature values must be entered in degrees Celsius.
•
Press the
Calibrate Gas Gauge
button to calibrate the temperature measurement system.
•
Deselect the
Calibrate
boxes after temperature calibration has completed.