Quick Start Guide
8
SBAU290 – September 2017
Copyright © 2017, Texas Instruments Incorporated
ADS54J64 Evaluation Module
6.
Press the
Additional Device Parameters
symbol next to the ADC Output Data Rate (see
), and
enter the following values, then press OK:
•
Check the
Enable
box.
•
ADC Sampling Rate = 983.04M
•
ADC Input Frequency = 90M
•
NCO = –122.88M
•
Decimation = 4
7. Press the CPU_RESET button on the TSW14J56EVM device.
8. Click the
Instrument Options
menu at the top of HSDC Pro, and select
Reset Board
.
9. Click
Capture
in HSDC Pro to capture data from the ADC.
10. In HSDC Pro, change the
Test Selection
to
Single Tone
.
11. Also in HSDC Pro, change the spectrum analysis from
Real FFT
to
Complex FFT
.
12. From the
Test Options
file menu, enable the
Analysis Window Markers
.
13. Set the value of the
Analysis Window Markers
in the lower left corner of the HSDC Pro window:
•
AWM1 = –120M
•
AWM2 = 120M
14. The results from the captured data of Channel 1 should resemble
, and the performance
should be similar to
. If this result was not achieved, then see
of this document.
Figure 5. Data Capture Results From Quick Start Procedure
Table 1. Quick Start Performance Measurements
Result
Measured Value
Units
SNR
> 67
dBFS
SFDR
> 80
dBFS