Quick Start Guide

2.4

Quick Start Trouble Shooting

Use

Table 2

to assist with problems that may have occurred during the quick start procedure.

Table 2. Troubleshooting Tips

Issue

Troubleshooting Tips

General Problems

Verify the test setup shown in

Figure 2

and repeat the setup procedure as described in this

document.

Check power supplies to the EVMs. Verify that the power switches are in the ON position.

Check signal and clock connections to the EVM.

Check that all boards are properly connected together.

Try pressing the

CPU_RESET

button on the TSW14J56EVM.

Try power-cycling the external power supply to the EVM and reprogram the LMK and ADC
devices.

TSW14J56 LEDs are not correct:

Verify the settings of the configuration switches on the TSW14J56EVM.

D1, D5 – N/A

Verify that the EVM configuration GUI is communicating with the USB and that the

D2, D4 – Blinking

configuration procedure has been followed.

D3, D6, D7 – OFF

(LEDs Not Blinking) Reprogram the LMK device.

D8, D28 – ON

Try pressing the

CPU_RESET

button on the TSW14J56EVM.

Try capturing data in HSDC Pro to force an LED status update.

Device GUI is not working properly

Verify that the USB cable is plugged into the EVM and the PC.

Check the computer’s Device Manager and verify that a

USB Serial Device

is recognized

when the EVM is connected to the PC.

Verify that the green

USB Status

LED light in the top right corner of the GUI is lit. If it is not

lit, press

Reconnect FTDI

button.

Try restarting the configuration GUI.

Check default jumper connections as shown in

Appendix A

.

HSDC Pro Software is not

Verify that the TSW14J56EVM is properly connected to the PC with an USB cable and that

capturing good data or analysis

the board serial number is properly identified by the HSDC Pro software.

results are incorrect.

Check that the proper ADC device is selected. In default conditions, ADS54J40_LMF_8224
should be selected.

Check that the analysis parameters are properly configured.

Check that the fundamental power is no larger than –1 dBFs.

HSDC Pro Software gives a Time-

Try to reprogram the LMK device and reset the JESD204 link.

Out error when capturing data

Verify that the ADC sampling rate is correct in the HSDC Pro software.

Sub-Optimal Measured

Make sure an ADC hardware reset was issued after loading the LMK but before loading the

Performance

ADC configuration file.

Check that the spectral analysis parameters are properly configured.

Verify that bandpass filters are used in the clock and input signal paths and that low-noise
signal sources are used.

9

SLAU652A – October 2015 – Revised January 2016

ADS54J40EVM

Submit Documentation Feedback

Copyright © 2015–2016, Texas Instruments Incorporated

Summary of Contents for ADS54J40

Page 1: ...3 HSDC Pro Settings 10 4 Software Description 11 4 1 ADS54J40 EVM GUI 11 4 2 Low Level View 12 5 Alternate Hardware Configurations 13 5 1 Clocking Options 13 5 2 Analog Input Options 14 Appendix A Ju...

Page 2: ...HSDC Pro software tool for high speed data converter evaluation The ADS54J40EVM was also designed to work with many of the development kits from leading FPGA vendors that contain an FMC connector 1 1...

Page 3: ...wer 1 3 Evaluation Board Feature Identification Summary The EVM features are labeled in Figure 1 Figure 1 EVM Feature Locations 1 4 References ADS54J40EVM software available at www ti com tool ADS54J4...

Page 4: ...t contains links to find the software on the TI website 2 Extract the files from the zip file 3 Run setup exe and follow the installation prompts 2 1 2 High Speed Data Converter Pro GUI Installation H...

Page 5: ...cedure The rest of this section describes the hardware setup steps Figure 2 Quick Start Test Setup 2 2 1 TSW14J56EVM Setup First setup the TSW14J56EVM using the following steps 1 Connect the ADS54J40E...

Page 6: ...Programs Texas Instruments ADCs ADS54Jxx EVM GUI 2 Verify that the green USB Status indicator is lit in the top right corner of the GUI If it is not lit click the Reconnect USB button and check the U...

Page 7: ...Figure 4 HSDC Pro GUI Main Panel 2 When prompted to select the capture board select the TSW14J56 whose serial number corresponds to the serial number on the TSW14J56EVM and click OK This popup can be...

Page 8: ...te and other parameters from the loaded configuration files Click OK 9 Click the Instrument Options menu at the top of HSDC Pro and select Reset Board 10 Click Capture in HSDC Pro to capture data from...

Page 9: ...d verify that a USB Serial Device is recognized when the EVM is connected to the PC Verify that the green USB Status LED light in the top right corner of the GUI is lit If it is not lit press Reconnec...

Page 10: ...ter SNR results The clock and analog inputs must be frequency locked such as through 10 MHz references in order to achieve coherency 3 3 HSDC Pro Settings HSDC Pro has some settings that can help impr...

Page 11: ...Low Level View tab to load configuration files LMK04828 Enables control of many of the LMK04828 features Configuration files can be used to setup the LMK04828 in known working configurations however...

Page 12: ...ll registers shown in the Register Map with the values shown in the Register Map log The log can be viewed by double left clicking in the bottom left status bar of this page Read Register button Read...

Page 13: ...YSREF signals to both boards This option provides the best performance as long as the clock source has better phase noise performance than the LMK04828 The source of the EXT ADC clock must be synchron...

Page 14: ...ired frequencies To use this mode load one of the configuration files named LMK_Config_Onboard_xxxx_MSPS cfg where xxxx corresponds to the desired ADC sampling rate A 10 MHz signal can be brought into...

Page 15: ...FMC Default is diff Shunt pins 2 3 A 2 Connector Descriptions The EVM connectors and their function are described in Table 7 Table 7 Connector Descriptions Connector Description J2 Channel A positive...

Page 16: ...ired Hardware section by replacing the reference to a 5 VDC power supply to power supply cable 2 Changed first bullet in the not included items list in the Required Hardware section to two 5 V power s...

Page 17: ...ing the warranty period to the address designated by TI and that are determined by TI not to conform to such warranty If TI elects to repair or replace such EVM TI shall have a reasonable time to repa...

Page 18: ...transmitter has been approved by Industry Canada to operate with the antenna types listed in the user guide with the maximum permissible gain and required antenna impedance for each antenna type indic...

Page 19: ...ified allowable ranges some circuit components may have elevated case temperatures These components include but are not limited to linear regulators switching transistors pass transistors current sens...

Page 20: ...REMOVAL OR REINSTALLATION ANCILLARY COSTS TO THE PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES RETESTING OUTSIDE COMPUTER TIME LABOR COSTS LOSS OF GOODWILL LOSS OF PROFITS LOSS OF SAVINGS LOSS OF USE L...

Page 21: ...sponsible for compliance with all legal regulatory and safety related requirements concerning its products and any use of TI components in its applications notwithstanding any applications related inf...

Reviews: