When the
Collect Data
operation is complete, the data displays the calculated channel statistics and a plot of
the data in the chart window. Various icons for viewing the data include zoom, pan, and home. The home icon
restores the graph to show all data collected. The viewing icons are underneath the statistics information on the
right side of the GUI.
4.2.1 Time Domain and Histogram Statistics
The
Time Domain
and
Histogram
plots share the same statistical information shown in
:
• Input
Channels
selected
•
Min
code within the data set
•
Max
code within the data set
•
Mean
code value within the data set
•
Std Dev
representing the standard deviation within the data set
•
Pk-to-Pk
representing the total noise peak-to-peak within the data set
•
Eff. Res
representing the effective resolution as number of bits with the value in parenthesis showing the
noise-free number of bits
Figure 4-7. Capture Statistics
4.2.2 Time Domain Plot
The
Time Domain
plot displays the sample count number in the X-axis. The
Y-axis
is displayed as either codes
or volts with the selection in a drop-down menu. The
Y-axis
drop-down menu displays next to the charting icons
near the right side of the GUI below the statistics.
shows the
Time Domain
plot and its features.
Figure 4-8. Time Domain Plot
ADS1x48EVM GUI
32
ADS1x48EVM Evaluation Module
SBAU378A – SEPTEMBER 2021 – REVISED JANUARY 2022
Copyright © 2022 Texas Instruments Incorporated