ADS131E08 Analysis Tools
25
SBAU200C – June 2012 – Revised April 2020
Copyright © 2012–2020, Texas Instruments Incorporated
Performance Demonstration Kit for the ADS131E08
5.3
FFT Tool
The
FFT
subtab displays the data in the frequency domain by performing an FFT on the eight channels.
Details of the FFT, including SNR, THD, and more, can be viewed using the
FFT Analysis
button located
at the bottom of the display.
shows the FFT display.
Figure 27. Analysis
→
FFT Graph of Normal Electrode Configuration
5.3.1
Coherent Frequency Calculator
The red box labeled
1
in
shows the Coherent Frequency Calculator. Coherent sampling in an
FFT is defined in
:
where:
•
f
AIN
= input frequency
•
f
SAMPLE
= ADS131E08 sampling frequency
•
N
WINDOW
= number of odd integer cycles during a given sampling period
•
N
TOTAL
= number of data points (in power of 2) that is used to create the FFT.
(3)
If the conditions for coherent sampling can be met, the FFT results for a periodic signal are optimized. The
ideal A
IN
frequency is a value that is calculated based on the sampling rate so that the coherent sampling
criteria can be met.