6 Using the product
30
5.
Confirm the entry with
[OK]
.
6.
Confirm the settings with
Apply
.
Country settings…
Instrument language and temperature unit can be set.
1.
[OK]
|
Configuration…
|
[OK]
.|
Country settings…
|
[OK]
.
- The
Country settings
dialogue is opened.
2. Move
Joystick
up/down to select the desired option.
- The selected option has an orange border ( ).
3.
Confirm the selection with
[OK]
.
- The selected setting is highlighted orange ( ).
4. Move
Joystick
up/down to change the setting.
5.
Confirm the entry with
[OK]
.
6.
Confirm the settings with
Apply
.
Optics...
The lenses adjusted to the instrument are shown. Only the lenses
that are shown may be used.
The high-temperature filter registered in the instrument is shown.
With use of another high-temperature filter, the specified accuracy
is not ensured.
>
[OK]
|
Configuration…
|
[OK]
|
Optics…
|
[OK]
.
Using the
Protection glass
option, you can set whether an infrared
protection glass is used or not. Ensure the correct setting to
prevent corruption of the measurement results.
> Press
[OK]
to activate (
) or deactivate (
) the
Protection
glass
option.
- With the option activated, the protection glass symbol (
) is
shown in the bottom right of the display.
SuperResolution (optional)
SuperResolution is a technology to improve image quality. Each
time an image is recorded, a sequence of images is saved on the
thermal imager and used to calculate a higher-resolution image
with the aid of the PC software (no interpolation). The pixel count
increases by a factor of 4, while the geometric resolution (IFOV) is
improved by a factor of 1.6.
To use the function, the imager must be hand-held (no tripod).