![TESTO 623 Instruction Manual Download Page 9](http://html.mh-extra.com/html/testo/623/623_instruction-manual_1092122009.webp)
6 Using the product
Pos: 27 /TD/Produkt verwenden/testo623/testo 623 Grenzwerte einstellen @ 4\mod_1254828443945_79.doc @ 51445 @ 255
6.2.
Setting the limits
Temperature
When changing the temperature unit from °C to °F, the set limit
values are converted
Humidity
When changing the humidity unit (%, td, wb), the set limit values
are not converted. The limit values must be reset via the Setup
menu.
Pos: 28 /TD/Produkt verwenden/testo623/testo 623 Einstellungen vornehmen neu @ 4\mod_1248956306675_79.doc @ 46623 @ 2
3.
6.3.
Performing settings
1. Press and hold
[Set]
for 2 s.
> Display
flashes.
2. Select the required function with
[Set]
.
>
The adjustable parameter flashes.
3
Set values with
[
◄
]
or
[
►
]
and confirm entry with
[Set]
(for
settings, see following table).
To scroll rapidly forward and back, press and hold
[
◄
or
►
]
.
Press
[Esc]
to cancel settings without saving them.
Hold
[Set]
down for 2 s to save previously set values and
return to the current reading display.
If the entire Setup menu is run through, the settings made
are automatically stored.
If no key is pressed within 30 s, the screen reverts to the
current reading display automatically. No settings are
saved.
-
Instrument changes to the next display function