
19
7. Measuring
7.4 Measuring undiluted CO
Always carry out a COundiluted measurement prior to a
flue gas measurement because the values for O2, CO and
in the instrument are overwritten during a COundiluted
measurement.
1 Select
CO undil
in the main menu
➀
and confirm via
.
-
The pump starts automatically and the measured value for
undiluted CO (uCO) is shown
➁
.
2 End measurement via
.
-
The last reading is retained in the instrument until the next
COundiluted measurement.
7.5 Measuring ambient CO
Always carry out an ambient CO measurement before a
flue gas measurement since a value for CO, retained in
the analyser, will be written over when ambient CO is
measured.
Carry out zeroing in fresh air (preferably outdoors) prior to
measuring ambient CO.
1 Select
CO amb
in the main menu
➀
and confirm selection
with
.
-
The pump starts automatically and the measured value for
ambient CO (
) is displayed.
2 End measurement via
.
-
The last reading is retained in the analyser until the next
ambient CO measurement.
➁
➀
➁
➀
CO undil
CO undil
CO amb
CO amb