TESTO 316-2 Instruction Manual Download Page 9

Maintaining the product

9

Maintaining the product

 Charging the rechargeable battery:

See chapter “First steps”

 Cleaning the sensor:

Tobacco smoke, dirty air, oils, grease, silicones and evaporating liquids or gases can 
lead to deposits on the surface of the sensor. This can result in reduced sensitivity 
and falsified display of concentrations. If necessary clean the sensor.
>  Switch on the instrument, wait until the end of the initialization phase and switch 

off again. Repeat this procedure several times.

 Clean the sensor head from the outside

>  If dirty, clean the sensor head with a soft dry cloth.

 Cleaning the housing:

>  If dirty, clean the housing with a damp cloth (soap solution). Do not use aggressive 

cleaning products or solvents!

 Switching on regularly

  If the instrument is used infrequently, deposits may build up on the sensor. 

Switching the instrument on prevents deposits building up on the sensor. Testo 
recommends switching the instrument on regularly to avoid deposits building up 
on the sensor.

 Storage & transportation

  To avoid contamination of the sensor, the instrument should not be stored 

or transported in an environment where there is tobacco smoke, foul air, oils, 
greases, silicones or evaporating liquids or gases. Any sensor that is contaminated 
as a result of storage or transportation must be cleaned before use, see Cleaning 
the sensor.

 Regular servicing:

Testo recommends yearly servicing of the gas detector by an authorized service 
centre. 

de

en

fr

es

it

pt

sv

nl

??

??

Summary of Contents for 316-2

Page 1: ...Instruction manual en testo 316 2 Leakage detector for gas...

Page 2: ...portant information Avoiding personal injury material damage Use the measuring instrument only for the purpose for which it is intended and within the parameters stated in the technical data Do not us...

Page 3: ...erature as well as the permitted operating temperature e g protect measuring instrument from direct sunlight Do not use testo 316 2 in closed rooms in which gases have collected into an explosive mixt...

Page 4: ...ry pack NiMh Battery life appprox 6h at 22 C 72 F Charging time approx 8h Dimensions 57 x 190 x 42mm Weight 348g Guidelines norms and tests Instrument complies with worksheet DVGW G465 4 Warranty Dura...

Page 5: ...hreshold Second alarm threshold Gas concentration maximum value display Maximum gas concentration detected since the last reset of the maximum display or since the last time the instrument was switche...

Page 6: ...8h In order to ensure as long a battery life as possible the battery shoud always be completely discharged and recharged The instrument can continue to be used during charging 1 Plug the mains plug i...

Page 7: ...r of burning from the hot sensor head after prolonged use Before touching the sensor head or packing up the instrument switch off and allow to cool Selecting gas to be detected Press several times unt...

Page 8: ...e g oils Do not use the instrument in dirty surroundings Move the sensor head as close as possible and slowly 3 to 5 cm per second over the parts which are to be examined for leaks Resetting maximum...

Page 9: ...ing with a damp cloth soap solution Do not use aggressive cleaning products or solvents Switching on regularly If the instrument is used infrequently deposits may build up on the sensor Switching the...

Page 10: ...Sensor connection faulty Steckverbidnung des Sensorkopfes pr fen Please contact your dealer or Testo customer service Sensor blinkt Sensor dirty clean sensor see chapter Maintaining product If we were...

Page 11: ...11 de en fr es it pt sv nl...

Page 12: ...0970 3165 en 04 0970 3165 en 05 0970 3165 en 05...

Reviews: