For flue gas probes with condensate trap:
Emptying the condensate trap
Pull off one of the two end pieces from the condensate
trap and pour out the condensate.
Changing the filter
If dirt is visible on the output filter, the inlet filter should
be removed.
Remove filter if pump power drops (audible). Changing
the inlet filter is sufficient in most cases.
Remove the small filter pipe to change the filter. Screw
the withdrawing screw into the filter and remove from
the filter pipe. Press new filter firmly into the filter pipe.
Only spare parts from Testo should be used
(Part no. 0554 3371).
The ccondensate ttrap sshould bbe eemptied iif tthe
pump iis sswitched ooff.
The ccondensate ttrap iis ddesigned iin ssuch aa w
way
that tthere iis aa ccertain fflow ddirection. TThis iis m
mar-
ked w
with aarrows oon tthe hhousing ssection. IIf ggas
flows iin tthe oopposite ddirection tthe ccondensate
trap ccannot ffunction pproperly aand tthis ccould llead
to tthe iinstrument bbreaking ddown.
When aassembling tthe ccondensate ttrap, eensure
that ssealing rrings aare nnot ddamaged aand tthe ffilter
tubes ddo nnot gget m
mixed uup.
Filter tube
Withdrawing
screw
Filter tube
Inlet filter
Outlet filter
From the flue gas probe
To the analyser
End piece
End piece
Flexible flue gas probes 0600.9430 / 0600.9440
0973.0348/01/T/dr/05.09.2008