LXATC-01 Infrared Body Temperature
Screening Camera
Technical Specifications:
Test Distance
1 meter
Lab Test Errors Range
33
°C to 37°C : error ±0.4°C
Below 33
°C, over 37°C : error ±0.6°C
Calibration
Computer software
Display Accuracy
0.1
°C
Alarm Temperature
37.4
°C or higher
Calibrate Temperature
36.5
°C with black body
Display Range
28
°C to 42°C
Ambient Temperature
16
°C to 28°C (indoor use, no wind)
Response Time
Within 2 seconds (standard mode)
Within 1.5 seconds (fast mode with PC)
Voltage Range and Display
DC10V-15V, DC12V/2A adapter included in
package. Lower than 10V LED shows dcL,
over 15V shows dcH.
Power Consumption
<24W
Alarm Way
Sound and flashing lights
Red highlight on PC screen
Working Temperature
-10
°C to 55°C
3