Test Products International Vib Meter 9085 User Manual Download Page 12

© TPI Europe 2021 

12                            TPI 9085 User Guide V1.0 

 

Factory Reset

 will reboot the 9085 and set all parameters back to factory defaults. 

 
The 

Graph Mode

 can be set to display the frequency spectrum as either a 

Line

 graph or 

Bar

 graph. 

 
The  display 

Color  Scheme 

can  be  configured  as  standard  (full 

colour)  or  monochrome,  e.g.  for  convenient  viewing  in  direct 
sunlight. 
 

Date & Time

 setting can be achieved using the up, down, left and 

right arrow buttons. 
 
Selecting 

Units 

from  the  Device  Settings  menu  allows  the  velocity  readings  to  be 

displayed in either mm/s or inch/s. Run Speed units can be displayed in units of Hertz

3

 

(Hz), revolutions per minute (RPM) or cycles per minute (CPM). The velocity type can be 
selected  between  RMS  and  Peak.  Displacement  units  can  be  set  to  Peak  or  Peak  to 
Peak. Temperature units can be set to Celsius or Fahrenheit.  
 
Selecting 

Temp Settings

 enables several temperature parameters to be set. 

Save with 

Vib

 can be enabled or disabled. When disabled the temperature reading will be displayed 

only and not saved for download. When enabled, the temperature reading will be saved 
with the vibration readings and transmitted to C-Trend software when the 9085 is docked. 

Save Mode

 can be set to Manual or Auto. In manual mode the 9085 will prompt you to 

press  the  center  key  after  the  countdown  timer  goes  to  zero  to  save  the  temperature 
reading. This can be useful if you would like to control when the reading is saved. In auto 
mode the temperature reading will be saved automatically after the countdown. Use 

Min 

Save  Time

  to  set  the  minimum  countdown  before  the  9085  prompts  you  to  save  the 

temperature  reading.  This  can  be  set  from  10s  to  120s. 

Auto  Save  Time

  allows 

adjustment of the time before the temperature reading is automatically saved. This can 
be set from 10s to 120s. 
 

Lock  Device

  immediately  locks  the  9085  to  prevent  unauthorized  use.  To  unlock  the 

9085  turn  the  9085  on  and  press  the  center  key  for  several  seconds  until  you  are 
prompted to enter the unlock code. The default unlock code is 375. 
 

Lock Settings

 locks out the ability to get into the menu to change settings without the 

unlock code. In this mode the 9085 will still take readings but the menu is inaccessible 
unless the unlock key is entered. To unlock and turn off this feature turn the 9085 on and 
press the center key for several seconds until you are prompted to enter the unlock code. 
The default unlock code is 375. 
 

Information

 displays the firmware version of the 9085 and the serial number. 

 

Temperature Only Mode

 displays only the temperature reading. Pressing and holding 

the menu key      puts the 9085 in temperature display only mode. Pressing the center 
test key      will return the 9085 to normal operation. 

 

3

 Hertz are equivalent to, and also sometimes referred to as, cycles per second (CPS). 

Summary of Contents for Vib Meter 9085

Page 1: ...er Guide V1 0 Vib Meter 9085 User Guide Version 1 0 Vibration Monitoring Analysis Test Products International 9615 SW Allen Blvd Suite 104 Beaverton OR 97005 USA Tel 800 368 5719 Fax 503 520 1225 Web...

Page 2: ...ASSET MANAGEMENT 13 4 C TREND II TRENDING SOFTWARE 14 5 SPECIFICATIONS 15 6 BASIC OPERATION 16 Disclaimer This document has been carefully prepared and checked No responsibility can be assumed for in...

Page 3: ...red analysed and trended with the included VibTrend PC analysis software 1 1 Special Requirement for Temperature Measurements To measure temperature with the 9085 the accelerometer with built in tempe...

Page 4: ...85 from the protective boot by prying the boot off from the bottom Set the 9085 on the charging pad so the symbol comes in contact with the pad The display of the 9085 will turn on to indicate it is c...

Page 5: ...cle button to turn the unit ON Place the sensor on a rigid part of the machine as close as possible to the desired measurement point e g bearing block using the magnet mount Pressing the circle button...

Page 6: ...ean ISO value mm s The ISO value in mm s or inch s is the large number displayed at the top of the screen which is the RMS average of the vibration velocity in the frequency band 10Hz 600 RPM to 1kHz...

Page 7: ...This reading is shown in units of g Earth s gravitational constant where 1g 9 81 m s2 Displacement Pressing either the left or right arrow button when the reading screen is displayed will cause the me...

Page 8: ...ignment This is based on the fact that shaft misalignment can result in a double peak in the waveform due to there being two different centres of gravity one from each shaft In other words the acceler...

Page 9: ...s It is possible to use live update with the basic readings screen the VA bands screen or 100 line frequency spectrum Pressing the circle button at any time during live update will capture a full 800...

Page 10: ...the ISO alarm levels to be set accordingly as does specifying the type of mounting rigid or flexible As a basic rule of thumb unless a machine is bolted down to a concrete floor its mounting should b...

Page 11: ...2 3 3 3 Sensor mV g Accelerometer sensitivity in mV g can be manually set using the up and down arrow buttons to allow use of any desired sensitivity sensor within the range 1mV g to 1000mV g This ena...

Page 12: ...countdown timer goes to zero to save the temperature reading This can be useful if you would like to control when the reading is saved In auto mode the temperature reading will be saved automatically...

Page 13: ...ected using the circle button The various options for each measurement point MP are then displayed 1 Take Reading takes a new reading and saves it to this MP 2 Save Reading saves the previously taken...

Page 14: ...stalled go to the activation screen and email the Hardware Key that is displayed to tech testproductsintl com C Trend II is very intuitive and very easy to drive and its full functionality is describe...

Page 15: ...s or inch s Bearing noise in BDU bearing damage units Displacement mm microns inch Temperature F C Displayed Frequency Units Hertz Hz RPM or CPM Input range User selectable with accelerometer sensiti...

Page 16: ...alarm colour coded in Green Amber and Red Additional screens show temperature displacement mm or microns frequency spectrum with cursor and diagnostics for unbalance misalignment and looseness Power O...

Reviews: