„STAT“:

 

It signalizes status radon probe according to following chart:

 

Color 

Description 

Green blink 3x 

Radon probe has just been turned on.  

Green blink after 5s 

Radon probe measures and works correctly 

Yellow blink 3x 

Radon probe has just been turned off. 

Green

 

Yellow 

 

blink after 5s

 

Radon probe measures but troubles are occur. 

 especially low capacity of 

accumulator. Warnings and errors are shown in PC application.  

No light, No blinking 

Radon probe 

doesn´t measure or accumulator is empty o

r device is damaged. 

Charging process of accumulator is described in chapter 

Basic Maintenance/ 

Accumulator charging

 

  

Antenna installation:  

   Supplied  antenna  is  screwed  on  antenna  connector.  When  installing  antenna,  hold  antenna  by  knurled 
end.

 

 

Power supply: 

According to operation method the radon probe can be supplied: 

1)

 

By  internal  accumulator  for  portable  use 

  Radon  probe  includes  internal  accumulator  which  is 

able  to  ensure  autonomous  operation  of  probe  for  more  than  12  months  without  charging.  
Depends  on  frequency  of  wireless  data  downloading  from  probe  and  depends  on  climatic 
condition of probe use. Accumulator is charged  over USB port and over provided USB cable. The 
USB cable is possible to connect to PC or to delivered power adapter. Status  of accumulator and 

charging process are described in paragraph ´Basic Maintenance/Accumulator charging´ 

 

2)

 

By  mains  power  supply  230V/50Hz  for  stationary  use 

  Radon  probe  is  permanently  supplied  by 

delivered power adapter. Power adapter is connected to probe via provided USB cable. In case of 
blackout internal accumulator ensures UPS function. 

 

 

Configuration:  

Setting and configuration is different according to operation way of radon probe. 
If  the  probe  is  used  for  autonomous  measurement  and  data  downloading  via  USB  interface  then  the 

setting  and  configuration  are  realized  by  usbRADONview  application.  UsbRADONview  application,  drivers 
and  user  manual  with  detail  configuration  description  are  free  downloaded  on  website: 

http://www.tesla.cz/

. 

If  the  probe  is  used  for  wireless  measurement  and  wireless  data  downloading  then  setting  and 

configuration  of  probe  and  whole  measuring  system  is  realized  by  Central  Unit  connected  to  PC  and 
TERAview application.

 

Central Unit is not included with package of probe and it is sold and delivered extra; 

see

 

http://www.tesla.cz/

.  TERAview  application,  drivers  and  user  manual  with  detail  configuration 

description  are  free  downloaded  on  website: 

http://www.tesla.cz/

.  For  successful  probe  configuration  in 

measuring system is essential to know probe radio channel number (communication wireless channel) and 
P2P address (identification in wireless net). Both parameters are printed out on probe serial number plate. 
Probe  radio  channel  number  is  possible  to  change  by  TERAview  application  and  it  must  be  identical  to 
central unit radio channel number. P2P address is permanent and it can occur in one big wireless net only 
once. Only Central Unit P2P address can be identical to P2P address of other elements in network. 
 

6

 

Basic Maintenance 

 
Accumulator charging: 

   During portable use of radon probe is essential to monitor state of internal accumulator and recharge it if 
necessary. If accumulator is discharged  the probe automatically turns off.  The probe is switched on again 
powering USB port. 
   Current state of accumulator can be determined by these ways: 

1) 

By LED diode ´STAT´

 - If LED starts blinking in green-yellow color it indicates that system is working 

incorrectly and one of main case is low voltage of accumulator. (see paragraph 

Operation Manual / 

LED diode “STAT””

)  

Summary of Contents for TSR4

Page 1: ...cal Specifications Operation Manual v 1 2017 Table of Contents 1 Introduction 2 2 Description and Utilization 2 3 Scope of Delivery 5 4 Product Specification 6 5 Operating Instructions 6 6 Basic Maint...

Page 2: ...d product If you experience any problems with your product please contact us at TESLA Rubeska 215 1 190 00 Prague 9 Vysocany www tesla cz 1 Introduction This document describes technical specification...

Page 3: ...battery power portable radon measuring probe supports flexible placing options within monitored structures Accumulator will last for more than 1 year after full charging After switching on TSR4 immedi...

Page 4: ...on and on the basis of measured set concentration level value it sends command to actuator wireless actuator or actuator in central unit can be use which is hardwired with power relay Power relay swit...

Page 5: ...ords Quicker download data from probes if you download data after end of the measurement Possibility of various identification writing into internal memory of probe Compatibility TSR4 and TSR3 radon p...

Page 6: ...n 18 2 hours Records saving interval probe 1 255 minutes default 1 hour Results memory capacity in probe 4096 150 days of 1 hours records Powering internal rechargeble accumulator charging via USB Acc...

Page 7: ...way of radon probe If the probe is used for autonomous measurement and data downloading via USB interface then the setting and configuration are realized by usbRADONview application UsbRADONview appl...

Page 8: ...n of Conformity will be delivered by Tesla producer on request If interested please use contacts on the web www tesla cz 8 Repairs Any repairs and non basic maintenance must be performed exclusively b...

Reviews: