background image

 

5/6 

5

 

Operating Instructions 

 
Switching on and off: 

   The probe measures radon concentration autonomously only if the switch is in position „I“( switch on). 

The switching on is signalized by LED diode „STAT“ according  chart below.  

 If the switch is in position „0“ (switch off) the probe doesn´t measure radon concentration.  In switching 

off mode the probe only keeps running real time for correct date and time of records in case of switch it on 
again.  By  switching  off  the  probe  doesn´t  lose  previous  records  of  measurement.  The  switching  on  is 
signalized by LED diode „STAT“ according  chart below.   

Download data from probe over USB is possible only if switch is in position „I“( switch on).   

 

LED diode „STAT“: 

It signalizes status radon probe according to following chart:

 

Color 

Description 

Green blink 3x 

Radon probe has just been turned on.  

Green blink after 5s 

Radon probe measures and works correctly 

Yellow blink 3x 

Radon probe has just been turned off. 

Green

 

Yellow 

 

blink after 5s

 

Radon probe measures but troubles are occur. – especially low capacity of 
accumulator. Warnings and errors are shown in PC application.  

No light, No blinking 

Radon probe doesn´t measure or accumulator is empty or device is damaged. 
Charging process of accumulator is described in chapter „Basic Maintenance/ 
Accumulator charging“ 

  

Power supply: 

According to operation method the radon probe can be supplied: 

1)

 

By  internal  accumulator  for  portable  use  –  Radon  probe  includes  internal  accumulator  which  is 

able  to  ensure  autonomous  operation  of  probe  for  more  than  12  months  without  charging.  
Depends on climatic condition of probe use. Accumulator is charged with USB port and provided 
USB  cable.  The  USB  cable  is  possible  to  connect  to  PC  or  to  delivered  power  adapter.  Status  of 
accumulator  and  charging  process  are  described  in  paragraph  ´Basic  Maintenance/Accumulator 
charging´ 

 

2)

 

By  mains  power  supply  230V/50Hz  for  stationary  use  –  Radon  probe  is  permanently  supplied  by 

delivered power adapter. Power adapter is connected to probe via provided USB cable. In case of 
blackout internal accumulator ensures UPS function. 

 

 

Configuration:  

Setting  and  configuration  are  realized  by  usbRADONview  application.  UsbRADONview  application, 

drivers  and  user  manual  with  detail  configuration  description  are  free  downloaded  on  website: 

http://www.tesla.cz/en/ke-stazeni/

. 

 
 

6

 

Basic Maintenance 

 
Accumulator charging: 

   During portable use of radon probe is essential to monitor state of internal accumulator and recharge it if 
necessary. If accumulator is discharged  the probe automatically turns off.  The probe is switched on again 
powering  USB port. 
   Current state of accumulator can be determined in two ways: 

1) By LED diode ´STAT´ - If LED starts blinking in green-yellow color it indicates that system is working 
incorrectly and one of main case is low voltage of accumulator. (see paragraph “Operation Manual / 
LED diode “STAT””)

.

 

2) In TERAview and usbRADONview application - where you can check current accumulator voltage. 
Voltage should not fall below 3.5 V, in limit conditions falls below 3.3V. 

Accumulator  is  charged  via  USB  port  using  supplied  USB  cable.  USB  cable  can  be  connected  to  PC  or  to 
supplied  power  adapter.  Connect  USB  cable  with  power  to  USB  port  of  probe.  LED  diode  ´CHRG´  next  to 
USB port of probe indicates charging status according to following chart: 