background image

 

 
 

 

 

 

 

 

 

 

 

 

 

 

 
 
 
 
 
 
 
 
 
 

Blank Page 

 

 

 

 

 

深圳德标仪器 

135-1095-0799

Summary of Contents for TC-5062C

Page 1: ...Manual R121114 http www tescom co kr Information contained in this document is subject to change without notice Copyright 2012 Tescom Co Ltd Suite 927 Unitechvil 142 Ilsan ro Ilsandong gu Goyang si Gyunggi do Korea 深圳德标仪器 135 1095 0799 ...

Page 2: ...Blank Page 深圳德标仪器 135 1095 0799 ...

Page 3: ...DUT Rotator Component Identification 13 1 7 Storage 14 Theory of Operation 15 2 1 Theory of operation 16 2 1 1 Standard E field Generation Ref 1 Figure 3 16 2 1 2 Radiation Power Measurement Ref 2 Figure 4 18 Maintenance 19 3 1 Maintenance of TC 5062C 20 3 1 1 Cleaning 20 3 1 2 Check List for Maintenance 20 3 2 Performance Test 22 3 2 1 Calibration Period 22 3 2 2 Required Equipment 22 3 2 3 Test ...

Page 4: ... Field Pattern 17 Figure 4 Radiation Power Measurement 18 Figure 5 TC 5062C RF Gasket 21 Figure 6 TC 5062C Shielding Measurement Test Set Up 22 Figure 7 Spectrum Analyzer Display 23 Figure 8 Typical VSWR of TC 5062C 24 Table of Contents Table 1 TC 5062C Connectors 11 Table 2 TC 5062C Optional I O Panel 12 Table 3 TC 5062C Accessories list 12 深圳德标仪器 135 1095 0799 ...

Page 5: ...tside of Korea TESCOM shall pay shipping charge to return the product to customer This warranty shall not apply to consumable parts and any failure or damage caused by improper use or unauthorized service In such cases TESCOM may refuse to furnish service under the warranty LIMITATION OF WARRANTY The foregoing warranty shall nor apply to defects resulting from improper or inadequate malignance by ...

Page 6: ...ualified service personnel Environmental Conditions Refrain from using this equipment in a place subject to much vibration direct sunlight and where the surface is not level Also use it where the temperature is between 0 C to 50 C and relative humidity is less than 85 Safety Symbols and Terms Terms in this manual WARNING Identifies conditions or practices that could result in injury or loss of lif...

Page 7: ...Any adjustment or repair of this product must be performed by qualified personnel Contact Information Address TESCOM Company Limited 927 Unitechvil 142 Ilsan ro Ilsandong gu Goyang si Gyeonggi do Korea 410 722 TEL 82 31 920 6600 FAX 82 31 920 6607 Email tescom sales tescom org Internet http www tescom co kr 深圳德标仪器 135 1095 0799 ...

Page 8: ...Service and Support 8 Blank Page 深圳德标仪器 135 1095 0799 ...

Page 9: ...1 General Information 9 This section provides a general description of the 6 GHz TEM cell and proper set up procedures General Information Chapter 1 深圳德标仪器 135 1095 0799 ...

Page 10: ...ange Key Features Reliable High RF Shielding up to 6 GHz Designed for long life Specifically designed for various types of mobile phones Absorbing materials for stable measurement results EMI filters on all data ports 1 2 Specifications General Specification VSWR 1 7 100 MHz 6 GHz Effective Cell Height 220 mm Field Strength at Center of Cell 13 dB V meter at 1 V input RF Connectors without module ...

Page 11: ...N Coaxial Connector Impedance 50 ohm Voltage Rating 1500 Vpeak Dielectric Withstanding Voltage 2500 Vrms SMA Coaxial Connector Impedance 50 ohm Voltage Rating 250 Vpeak Dielectric Withstanding Voltage 750 Vrms 1 4 Component Identification N f RF Connector SMA f RF Connector M506202A DB module Door Handle Door Figure 1 TC 5062C Component Identification 深圳德标仪器 135 1095 0799 ...

Page 12: ...DB25 Data Interface Panel Shielding Spec 70 dB from 0 1 to 2 GHz 60 dB from 2 to 3 GHz 55 dB from 3 to 6 GHz Working Voltage 100 VDC Dielectric Withstanding Voltage 300 VDC EMI Filter 1000 pF Pi filter M506204A USB2 0 Data Interface Panel Shielding Spec 60 dB from 0 1 to 6 GHz USB A 2 0 outside and inside Table 3 TC 5062C Accessories list Part Number Name Spec Quantity Operating Manual Test Report...

Page 13: ...red at different direction by rotating the rotator handle Stick for the DUT position Control Measurement result is depend on the position of DUT antenna To get stable result the DUT should be properly positioned using this stick The stick must be removed before install the DUT Rotator CAUTION To avoid hazardous electrical shock do not perform electrical tests when there are signs of shipping damag...

Page 14: ...The storage temperature range for this equipment is 20 C to 70 C When this equipment is not used for a long period of time cover with vinyl or place in a cardboard box and store in a dry place away from direct sunlight 深圳德标仪器 135 1095 0799 ...

Page 15: ...This section provides important information regarding TC 5062C operation Operation Chapter 2 深圳德标仪器 135 1095 0799 ...

Page 16: ...ssentially the same as with the TEM Cell The E H field inside the test volume is proportional to the input voltage and inversely proportional to the cell height If a radiating object is inserted inside the cell the radiated wave toward the input port is guided by the transmission line and picked up at the input with a receiver such as a spectrum analyzer With this method RFI from a radiating Devic...

Page 17: ...n 100 MHz to 6 GHz and the signal level to 220 mV The E field developed at the test point inside TC 5062C is then E 0 22 V 0 22 meter 1 Volt meter If the signal generator output is decreased to 2 2 uV the E field inside TC 5062C becomes 10 uV meter at the center of the test volume L meter V volt E volt meter DUT E b Field Pattern Top View a Field Generation Signal Generator 深圳德标仪器 135 1095 0799 ...

Page 18: ...ct is in a free space and has antenna gain G We also assume that in our measurement the radiator size is sufficiently small compared to the TEM cell such that there is no significant effect on radiation due to the TEM Cell Then the hypothetical maximum EMI field Er at distance D meter from the radiator would be Er Volt meter 60 x G Zo x Vm D 5 Vm is the maximum measured voltage which is the vector...

Page 19: ...This section contains information for keeping the instrument in good working order and checking its overall performance Maintenance Chapter 3 深圳德标仪器 135 1095 0799 ...

Page 20: ...black oxide build up 3 1 1 Cleaning Periodically wipe the case with a damp cloth mild detergent do not use abrasives or solvents 3 1 2 Check List for Maintenance 1 Check for loose screws and tighten with proper tools if necessary 2 Check for a loose connection If a loose connector is found tighten the connector with proper tools 3 Check for a damaged cable especially near the connector cable neck ...

Page 21: ...hey will damage the plastic parts Use alcohol to clean and maintain parts of the equipment The RF gasket is an important element of optimum shielding It must be checked periodically for damage or wear Periodic cleaning of the aluminum contact surface where the gasket touches is important for optimum shielding performance 深圳德标仪器 135 1095 0799 ...

Page 22: ... 6 GHz TEM Cell specifications every time maintenance is performed on a 6 GHz TEM cell such as replacing the gasket during routine maintenance Also performance tests the 6 GHz TEM Cell if you suspect mechanical damage 3 2 1 Calibration Period Shielding Effectiveness 6 months VSWR and Insertion Loss 1 year 3 2 2 Required Equipment Spectrum Analyzer 6 GHz Network Analyzer 6 GHz Signal Generator 6 GH...

Page 23: ...enna to the Spectrum Analyzer with RF cable 5 Open the door of the TC 5062C and move the antenna around the TC 5062C to find the location where the maximum field is found Fix the location of antenna for maximum field 6 Adjust the spectrum analyzer input range and set the signal level to the top display line 7 Close the lid and probe around the lid gasket seams RF connections and cable joints Look ...

Page 24: ...p 6000 MHz Measure S11 Format SWR 2 Calibrate the Network Analyzer in order to measure S11 3 Connect the calibrated cable to the input port of TC 5062C and measure SWR 4 Verify S11 SWR of 5062C meets its specification 3 2 4 Typical VSWR of TC 5062C VSWR 30 KHz to 6000 MHz Scale 1 Div Figure 8 Typical VSWR of TC 5062C 深圳德标仪器 135 1095 0799 ...

Reviews: