79
– This action is confirmed by extinction of the
sign and the display of END.
– Switch off the instrument and change the selector 3 on position measurement.
– Switch on the instrument and realise measurement.
If the calibration values are incoherent, an error code appears and the values are not taken
into consideration.
After the instrument calibration and the setting of the measuring device, you can deactivate
the functions keys, this protects the operator against faulty input and faulty measuring
results. Only the key
, the external data switch entry and the R-M command are still
active, allowing measurements.
To deactivate the keyboard, switch off the instrument after setting the measuring device,
switch the selector 4 to ON, power on again the instrument. The lock sign appears on the
display, the keyboard is deactivated.
To animate the keyboard again, practice as before but switch the selector 4 to OFF.
M
9.5 Resolution of numerical display
The maximum resolution of the instrument is DP
= 0.1 µm / 5 µin.
For measurements not requiring this resolution the resolution
DP
= 1 µm / 50 µin is proposed.
Example :
DP
706
µm
.02780
in
DP
705.8 µm
.027785 in
9.6 Selection of number of classes (TT60)
Number of classes within tolerance
SW6
SW7
5
OFF
OFF
10
ON
OFF
20
OFF
ON
40
ON
ON