6
3. SPECIFICATIONS
3-1 General specifications
•
Measurement method: Digital, triaxial measurement.
•
Directional characteristic: Isotropic, triaxial.
•
Measurement range selection: One continuous range.
•
Display resolution: 0.1mV/m, 0.1
µ
A/m, 0.1
µ
W/m
2
, 0.001
µ
W/cm
2
•
Setting time: Typically 1s (0 to 90% of measurement value).
•
Display refresh rate: Typically 0.5 seconds
•
Display type: Liquid-crystal (LCD), 4 digit.
•
Audible alarm: Buzzer.
•
Units: mV/m, V/m,
µ
A/m, mA/m,
µ
W/m
2
, mW/m
2
, W/m
2
,
µ
W/cm
2
, mW/cm
2
•
Display value: Instantaneous measured value, maximum
value or maximum average value.
•
Alarm function: Adjustable threshold with ON/OFF.
•
Calibration factor CAL: Adjustable.
•
Manual data memory and read storage: 99 data sets.
•
Rechargeable lithium-ion battery: 3.6V, 1940mAh
•
Battery charging time: Approx. 3 hours
•
Battery operation time: Approx. 10 hours
•
Auto power off: 5 minutes.
•
Operating temperature range: 0
℃
to +50
℃
•
Operating humidity range: 25% to 75%RH
•
Storage temperature range: -10
℃
to +60
℃
•
Storage humidity range: 0% to 80%RH
•
Dimensions: Approx. 67(W)
×
60(T)
×
247(L)mm.
•
Weight (including battery): Approx. 250g
•
Accessories: Instruction manual, Rechargeable Lithium-ion
battery, Carrying case, AC adaptor (DC 5V 1.5A).