Table 17-3: Physical specifications (continued)
Characteristic
Specification
Gas supply requirements
• Connector location: when facing the back of the device, the connector is located on
the back on the bottom
‑
left corner.
• Inlet pressure range: regulated between 40 psi and 60 psi.
Set point pressure is 45 +15/−5 psi.
• Oxygen volume in the gas supplied must not exceed 20%.
• Device includes an external pressure relief valve rated to 75 psi.
– Discharge area: when facing the back of the device, the discharge area is
located on the back on the bottom
‑
left side.
• Internal: 40
‑
micron filter installed.
• Device has a 303
‑
stainless steel, ARO-shape hose coupling plug.
Gas usage
One gas tank with a volume of 6.26 m³ (221 cu. ft.) lasts approximately 30 days under
normal operating conditions.
Touch Screen Specifications
Table 17-4: Touch Screen Specifications
Characteristic
Specification
Touch screen display
• Width: 30.5 cm (12 in)
• Height: 22.9 cm (9 in)
• VGA liquid crystal display (LCD)
System Specifications
17-4
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...