Installing the Quantum System
Warning:
Only trained and qualified personnel should move the Quantum device.
Caution:
Do not block access to the primary power switch, which is located on the back of the device, during
installation.
Installation information is included in the shipping container of each Quantum system. An authorized
service representative uses this information to install the system in your facility. The service
representative also tests the system to confirm that it is functioning properly before you begin using it.
Electrical Requirements
Warning:
To reduce the risk of electrical shock, do not use an adapter that breaks the protective
ground.
Connect the Quantum system power cord only to a properly installed, three
‑
wire
‑
grounded,
hospital
‑
approved grounding
‑
type electrical receptacle. See, “Electrical Power and Safety
Specifications” on page 17-2 for the electrical specifications for the Quantum device.
Maintaining the Quantum System
Caution:
Do not lubricate the pumps or the pump rotors for any reason.
Follow the instructions in this chapter to clean and disinfect the Quantum system. Other than to ensure
that the system is clean and undamaged before use, you are not required to maintain the system.
Cleaning the System
Warning:
Turn off the primary power switch and unplug the device before cleaning.
Warning:
Use a cleaning technique that does not place your finger in a vulnerable position. For
example, you can clean the area around the valve with a cotton swab or by using a flossing technique.
Warning:
Take adequate precautions when cleaning any surfaces of the device that might have been
exposed to blood to prevent possible exposure to and transmission of infectious diseases.
When cleaning the Quantum system, remove dirt and fluid using a soft cloth and a solution of water and
a mild detergent that is safe for plastic. To avoid scratching the touch screen, do not clean the touch
screen with an abrasive brush or scrub material. Do not remove the pump rotors or the mounting plate
when cleaning the system. To disinfect the system in the case of a biohazardous spill, follow the
instructions in “Disinfecting the System.”
Maintaining the Quantum System
16-2
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...