Harvest Cells
The purpose of this task is to transfer cells in suspension from the IC circulation loop, including cells in
the bioreactor, to the harvest bag.
Table 12-12 describes the types of solution that are needed to attach to each line when performing
Harvest Cells. These solutions and corresponding volumes are based on the factory default settings for
this task.
Note:
When harvesting cells, you should adjust the media composition in the harvest bag to ensure that it is
appropriate for the way in which you intend to use the cells after harvest.
Table 12-12: Solutions for Harvest Cells
Line
Solution
Attached to the
Line
Required Volume (estimate based on factory default settings)
Cell
None
N/A
Reagent
None
N/A
IC Media
Harvest media*
0.4 L
Wash
None
N/A
EC Media
Media without
protein
0.1 L
*When harvesting non-adherent cells, you can use PBS or a similar electrolyte solution as the harvest media; in this situation the harvest
media does not have to consist of a media formulation, because there is no trypsin to neutralize.
Harvest Cells pathway:
Task
>
Release and Harvest
>
Harvest Cells
1.
Confirm the settings for Harvest Cells shown in Table 12-13.
Table 12-13: Settings for Harvest Cells
Setting
Factory Default
Setting Options
IC Inlet
IC Media
Reagent, IC Media, Wash, EC Media
IC Inlet Rate
400 mL/min
100 to 500 mL/min
IC Circulation Rate
−70 mL/min
−87 to −18 mL/min
The system automatically calculates this rate based on the IC Inlet Rate
for this task.
EC Inlet
EC Media
Reagent, IC Media, Wash, EC Media
EC Inlet Rate
60 mL/min
0 to 100 mL/min
Harvest Cells
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
12-9
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...