Inlet Line Washout
The purpose of this task is to wash a line or to add fluid to a bag using fluid from another bag. For
example, following a bone marrow load, you could wash the droplets left in the cell line back into the cell
inlet bag, which would leave a clean line for future use.
Note:
During this task, the system opens the appropriate valves to provide flow between the selected bags. You
must determine both the direction and the force of the fluid flow by adjusting the relative height of the source and
destination bags.
1.
Touch
Task
.
2.
Touch
Washout
.
3.
Touch
Inlet Line Washout
.
The Setup: Inlet Line Washout screen appears, which prompts you to select the inlet source and the
inlet destination, as shown in Figure 9-1.
Figure 9-1: Setup: Inlet Line Washout screen
4.
Touch the Inlet Source field.
A menu of inlet source options appears, and the Inlet Source field changes to black. The available
inlet source options for this task include Cell, IC Media, Reagent, EC Media, and Wash.
5.
Select the desired inlet source from the menu.
The Inlet Source field displays the option you selected, and the Inlet Source field changes to gray.
Inlet Line Washout
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
9-9
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...