Remove EC Air
Caution:
Periodically inspect the EC (extracapillary) inlet port and the EC outlet port for air, and if either port
contains visible air, remove the air by performing the Remove EC Air task.
The purpose of this task is to remove any visible air from the inlet port and outlet port of the EC side of
the bioreactor. The system starts this task by removing air from the EC outlet, then it removes air from
the EC inlet, and then it removes any residual air left in the EC outlet. Figure 8-3 shows the EC inlet and
EC outlet ports located on the bioreactor.
Figure 8-3: EC ports
1
EC inlet port
2
EC outlet port
To perform Remove EC Air:
1.
Touch
Task
.
2.
Touch
System Management
.
3.
Touch
Remove EC Air
.
The Setup: Remove EC Air screen appears, which prompts you to select the inlet source, as shown
in Figure 8-4.
System Management Tasks
8-4
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...