Guidance and manufacturer’s declaration – electromagnetic immunity
The thermometer are intended for use in the eIectromagnetic enviroment specified
below. The customer or the user of these thermometer should assure that it is used
in such environment.
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV
contact ±8
kV air
±6 kV
contact ±8
kV air
Floor should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Electrical fast
transient/burst
IEC 61000-4-4
Not applicable Not applicable Not applicable
Surge IEC
61000-4-5
Not applicable Not applicable Not applicable
Voltage dips, short
interruptions and
voltage variations
on power supply
IEC 61000-4-11
Not applicable Not applicable Not applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.