Termite detection systems TDS-II Operator'S Manual Download Page 4

the unit into TURBO mode.  Pressing this button 4 times rapidly sets the TURBO mode and the 
LED lowest green light will appear to shimmer. 

 

 

LOW-BATT LED      

The LOW-BATT LED  will  become  lit  when there is  approximately  one  hour  left on  the  current 
battery charge.  During the charging process, this LED will blink off and on until the battery is 
fully charged; at that time the LED will turn off and it will signal a fully changed unit. 

(NOTE: If desired the TDS-II will fully operate when the battery is being charged) 

 

TERMITE DETECTION PROCEDURE 

 

The TDS-II electronics warm-up and self calibration takes 30 seconds.  Completion of this start 
up  process  is  signaled  by  a  rapid  flashing  of  the  bar  graph  LEDs  followed by  a  continuous  1 
second beep.  The unit will default to medium (M) sensitivity at start up.  The user may chose to 
change the sensitivity, MUTE or PEAK functions at this time or at any time during operation and 
the change will take place instantaneously. 

The  TDS-II  will  equalize  to  its  current  background  (or  ambient)  condition.  So  even  with  an 
elevation of  termites’  gases  in  the  surrounding air  it  will  only  detect on  areas  that  exceed this 
current background.  This will enable the user to detect the location of large concentrations and 
not trigger on false signals. 

 

Detection of Termites in Closed Spaces     

Often  large  concentrations  of  termites  can  be  found  in  wall  spaces.    Here  the  gases  they 
produce will be largely trapped and creating a different atmosphere than the surrounding area.  
This is the easiest type of termite infestation to detect.  Since the gases given off are heavier 
than  air  they  will  tend  to  sink  to  the  bottom  of  the  enclosed  area.    If  this  area  is  wall  space 
detection may be triggered by checking along base boards or in electrical outlet boxes.  Move 
the TDS-II probe tip slowly along a suspected area (2-3 inches per second).  If a signal occurs 
go back to the area and check closer.  Subsequent positive signals at suspected area strongly 
suggests the presents of termites. 

If closer inspection of the area is warranted, a small 1/8  – 5/32 inch hole can be drilled into the 
wall space for inspection.  Using the needle nose probe insert it in the hole for inspection of any 
termite gases.  If termites are detected an audio signal will sound and the graph bar LEDs will be 

Summary of Contents for TDS-II

Page 1: ......

Page 2: ...r user s convenience Three levels of sensitivity are provided to aid the user in isolation of small and large concentrations of termites The fourth setting the TURBO setting is super sensitive and is recommended for the detection of Bed Bugs Since the TDS II is self calibrating no special set up is required The detector is ready for use within 30 seconds after being turned on A standard 9 inch wan...

Page 3: ...e LED The TDS II starts in medium M level which is a sufficient sensitivity for normal colonies of termites commonly found in the subterranean species The high sensitivity H is reserved for more difficult detections of small swarms such as dry wood termites The low L sensitivity is used when large concentrations are present and the user is trying to isolate to specific areas of a concentration The...

Page 4: ...n elevation of termites gases in the surrounding air it will only detect on areas that exceed this current background This will enable the user to detect the location of large concentrations and not trigger on false signals Detection of Termites in Closed Spaces Often large concentrations of termites can be found in wall spaces Here the gases they produce will be largely trapped and creating a dif...

Page 5: ...then quickly press the PEAK button four times The TURBO mode is indicated when the lowest LED green light appears to shimmer Now the unit is in the highest sensitivity setting This setting is super sensitive and will easily alert to the small amount of CO2 produced by the bed bugs To test for bed bugs move the probe slowly near the areas being checked for bed bugs Do not block the probe tip If a b...

Page 6: ...itional validation to the user that the infestation is gone TDS II USE AND CARE Although the TDS II is stable it can be set off by excess turbulent air currents around its nozzle Due to this use of the TDS II outside may lead to mixed results and can result in false signals on the most sensitive levels The TDS II is a very sensitive instrument that can detect extremely small concentrations of CO2 ...

Page 7: ... is desired to be replaced the same wire may be used to push out the old filter A new filter must replace the old one Running the unit without an approved filter may damage the unit Battery Charging The battery supplied with the TDS II is a rechargeable lithium ion polymer battery that will run continuously for 8 eight hours before a charge is needed Recharging to a full state requires about 3 hou...

Page 8: ... off 10 minutes of inactivity Battery 3 7V 1800 mAh rechargeable Li ion polymer Compatible models include Fujifim NP120 Pentex DLi7 or Richoh DB43 Re charge time approximately 3 hours to full charge Operating environment 320 F 00 C to 1220 F 400 C at 75 RH Storage environment 80 RH for detector and battery storage temperatures are recommended to be the same as operating temperatures to maximize ba...

Page 9: ...and closer and slower to suspected area replace filter and or clean probe 2 TDS II will not turn on ON OFF button not held for 1 second to turn on Battery is discharged or dead Hold ON OFF button down for 1 second Recharge battery If it will not hold charge replace battery 3 The TDS II will trigger inadvertently Turbulent surrounding air can cause detector to trigger Large Attempt to avoid areas a...

Page 10: ...g detected signals the TDS II needs about 3 5 seconds to settle down Solution Pull TDS II away from the detected area for 3 5 seconds before repeating to that location again 5 The TDS II appears to be taking longer to detect than The wand tip or needle probe filter has become constricted due to moisture or dirt Replace the filter with a new one and discard the old one Check and clean needle probe ...

Page 11: ...the original purchaser against detects in material or workmanship for a period of one 1 year from the date of purchase by Termite Detection Systems Inc TDS During this warranty period TDS will at its option replace or repair the defective unit The warranty does not apply to defect resulting from abuse neglect accident unauthorized repair alternation or unreasonable use of the instrument Any implie...

Page 12: ...ole discretion of TDS For out of warranty service obtain an RMA number and send unit to the following address Termite detection systems Inc 333 N E 46th Street Oak Island N C 28465 Phone 910 448 1003 Email sales termitedetector com Rev 05 10 P N 950 001 Each TDS II Contains TDS II Infrared Termite and Bed Bug Detector 9 Flexible Extension Wand and 5 Needle Probe A C Wall Charger 12V Car Charger Di...

Reviews: