Terahertz Technologies TIA-1200 Operating Instructions Manual Download Page 4

Page 2

Operating Considerations

The TIA-1200 is comprised of a 

fi

 ber coupled InGaAs detector coupled with a reverse bias voltage 

network. The ouput of the diode is brought out to a type K female SMA connector. Light falling on the 
detector generates a positive-going proportional current. This current, multiplied by the load imped-
ance produces a voltage that is proportional to the light incident on the detector surface.

Normally the unit is used to drive a coaxial cable, this cable should have a 50 ohm characteristic 
impedance and be terminated with a 50 ohm load at the oscilloscope or other measuring device to be 
used. 

The active area of the TIA-1200 is 50 microns in diameter. With the optional 50 micron 

fi

 ber both sing-

lemode and multimode sources may be tested without incurring excessive coupling losses. 

Typical optical return loss is 35 dB and is caused primarily by the FC/PC connector employed. If 
higher return loss is needed, The unit may be optionally equipped with singlemode angled PC con-
nectors for ORLs of around 52 dB.

 Typical Setup of the TIA-1200

“K” Connector

FC Connector

Signal

Model TIA-1200 O/E Converter

Power

Off

On

Output 
50 

Ω

User supplied RF 
output cable - 50 Ohm

Ω

TTI

Summary of Contents for TIA-1200

Page 1: ...TIA 1200 Optical Electrical Converter Terahertz Technologies Inc 169 Clear Road Oriskany NY 13424 315 736 3642 FAX 315 736 4078 E mail sales terahertztechnologies com 5 2013 Operating Instructions...

Page 2: ...ents Introduction 1 Unpackaging and Inspection 1 Power Supply 1 Typical Setup 2 Operating Considerations 2 Typical Frequency Response 3 Spectral Response 3 Specifications 4 Service Warranty Informatio...

Page 3: ...chanical and electrical defects Upon acceptance by the carrier he assumes responsibility for its safe arrival After unpack ing examine the unit for any evidence of shipping damage Should you receive t...

Page 4: ...50 ohm characteristic impedance and be terminated with a 50 ohm load at the oscilloscope or other measuring device to be used The active area of the TIA 1200 is 50 microns in diameter With the option...

Page 5: ...Frequency GHz Relative Power Response dB Typical Frequency Response The approximate relative response curves of the detectors employed is as shown below Note that these are representative curves and...

Page 6: ...nput Connector FC or FC APC Input Numerical Aperture 0 11 Optical Return Loss 35dB 52dB with APC Connection Power Requirments 120 240 VAC 50 60Hz Dimensions 1 2 W x 2 5 L x 1 3 H 30 5mm x 63mm x 33mm...

Page 7: ...ODUCT I E TTI WILL NOT BE RESPONSIBLE FOR ANY PERSONAL INJURY PROPERTY DAMAGE OTHER THAN THE COST OF REPLACING THE PRODUCT OR ANY OTHER MONETARY DAMAGE SUCH AS LOST WAGES OR PROF ITS CAUSED BY ANY USE...