BTX-1 Series Instrument Revised: 3.14.2018
www.tensitron.com
pg. 11
▪
Scroll to the description to rename and press ENTER.
▪
Input your custom name by using the up and down arrows to
select each number, letter, or character, followed each time
by ENTER.
▪
Continue pressing ENTER until all spaces in the description
have values, including blank spaces.
▪
Once your new name has been entered you will automatically
be returned to the SETUP screen.
o
Re-Cal Tension: See description at the end of this booklet.
O
NLY PERFORM THIS IF YOU ARE A QUALIFIED CALIBRATION HOUSE USING
NIST-
CERTIFIED
CALIBRATION WEIGHTS
.
N
OT USED FOR CHECKING ACCURACY
.
T
O CHECK ACCURACY
,
REFER TO
S
EC
.
7:
C
HECK
A
CCURACY
.
o
Audio
To turn the Beep ON or OFF:
▪
Scroll to SETUP and press ENTER.
▪
Scroll to AUDIO and press ENTER.
▪
Select ON or OFF and press ENTER.
o
Backlight
To adjust the visual intensity of the LCD screen.
▪
Scroll to SETUP and press ENTER.
▪
Scroll to BACKLIGHT and press ENTER
▪
Select Low, Medium or Full intensity and press ENTER.
o
Set-Point Menu
This feature displays tension as a colored bar on the main display. The
colored bar advances as tension increases, and retracts as tension
decreases. You can define a safe operating tension range. For tension
measurements within this range, the bar is green. Below the range, the
bar is amber. Above the range, the bar is red.
To adjust the Set Point Menu:
▪
Scroll to SETUP and press ENTER.
▪
Scroll to SETPOINT MENU and press ENTER.
▪
Follow and enter the prompts to turn this feature on or off, or
to edit your high and low values. (See Figure 6)
Set Point Setup Screen
See High and Low threshold values
Settings:
On / Off
Edit High Point Threshold
Edit High Point Threshold
Figure 6: Setpoint Setup