ACX-1 Series Instrument
www.tensitron.com
pg. 11
6. DISPLAY SCREENS
•
SETUP Screen
o
LCD Refresh Rate
To either speed up or slow down the instrument’s LCD refresh rates:
▪
Scroll to SETUP and press ENTER.
▪
Scroll to LCD REFRESH RATE and press ENTER.
▪
Select your preference: 1, 2 or 5 Hz and press ENTER.
Note: This feature is unrelated to Data Averaging
o
Data Averaging
Note: This feature is primarily used in processes where the material to
be checked is constantly moving, such as coil winding, and textiles, and
likely does not apply to use on Aircraft cables.
o
Custom Names
Custom calibrations, listed as Custom 0
– 9, can be renamed so your
material appears as a selection on the Main Display. To rename a
Custom Calibration:
▪
Scroll to SETUP and press ENTER.
▪
Scroll to CUSTOM NAMES and press ENTER.
▪
Scroll to the description to rename and press ENTER.
▪
Input your custom name by using the up and down arrows to
select each number, letter, or character, followed each time
by ENTER.
▪
Continue pressing ENTER until all spaces in the description
have values, including blank spaces.
▪
Once your new name has been entered you will automatically
be returned to the SETUP screen.
o
Re-Cal Tension: See description at the end of this booklet.
O
NLY PERFORM THIS IF YOU ARE A QUALIFIED CALIBRATION HOUSE USING
NIST-
CERTIFIED
CALIBRATION WEIGHTS
.
N
OT USED FOR CHECKING ACCURACY
.
T
O CHECK ACCURACY
,
REFER TO
S
EC
.
7:
C
HECK
A
CCURACY
.
o
Audio
To turn the Beep ON or OFF:
▪
Scroll to SETUP and press ENTER.
▪
Scroll to AUDIO and press ENTER.
▪
Select ON or OFF and press ENTER.
o
Backlight
To adjust the visual intensity of the LCD screen.
▪
Scroll to SETUP and press ENTER.
▪
Scroll to BACKLIGHT and press ENTER
▪
Select Low, Medium or Full intensity and press ENTER.