30
20. GENERAL SPECIFICATION
Waveform
Asymmetrical rectangular
Amplitude (over 500 Ohm load) 99.5 mA +/- 10%
Max intensity
50V zero to peak across 500
Ω
Constant voltage over 470-1500 Ohm
Constant current over 160-470 Ohm
Output plug
Fully shielded
Channels
Single channel
Max Pulse energy
Total output limited to 25
μC per pulse
Batteries
2 x AA alkaline (two AA batteries) or 2 x AA NiMH
Weight
90 g
without batteries
Dimensions
102 x 53 x 28 mm
Safety Classification
Internal power source.
Environmental Specifications:
Operating:
Storage:
Temperature range: 0 to +40
⁰
C
Humidity: 10 to 93% RH non-condensing
Atmospheric pressure: 700hPa to 1060hPa
Temperature range: -10 to +60
⁰
C
Humidity: 8 to 93% RH non-condensing
Atmospheric pressure: 700hPa to 1060hPa
TYPE BF
APPLIED PART
Equipment providing a degree of protection against
electric shock, with isolated applied part.
Designed for continuous use.
This symbol on the unit means “Refer to Instructions
for Use”
IP22
The unit is not water resistant, and should be
protected from liquids.
Complies with EU WEEE regulations
Applied Part
Vaginal and anal electrodes. Optional skin surface
electrode pads. See section 17.
Contact duration: At least 10 minutes.
Note:
The electrical specifications are nominal and subject to variation from
the listed values due to normal production tolerances of at least 5%.